![YOKOGAWA 701924 User Manual Download Page 19](http://html.mh-extra.com/html/yokogawa/701924/701924_user-manual_911102019.webp)
7
IM 701924-01E
Probe-Head End
4.
Pass the lead wire through the heat-shrink tube.
* The heat-shrink tube for the probe-head end is not included. It must be
obtained separately.
5.
Solder the lead wire’s core wire to the flanged input pin.
6.
Cover the flanged input pin with a heat-shrink tube, and then apply heat
with a drier to fix the tube in place.
Do not cover the section of the flanged input pin to the left of the broken
lines shown in the following figure with the heat-shrink tube. Make sure
that the diameter of the tube after shrinking is 2.0 mm or less. If the
maximum diameter exceeds 2.0 mm, the lead wire will not be able to
pass through the retaining cover.
4.0 mm
(soldering
not allowed*)
φ2.0 mm or less
* If solder gets on this section (4 mm from the tip), the probe may break.
Flanged input pin
Cover with a heat-shrink tube
Solder
Core wire
Lead wire
Heat-shrink tube