background image

—  13  —

IM 3201–E

5. Maintenance

5.1 Adjustment

In case of the accuracy of the tester becomes doubtful as the 
result that an excessive load of shock is applied to the tester, 
the tester must be tested using other calibrated instrument. If the 

accuracy  is  observed  to  be  over  the  specified  value,  the  tester 

must be readjusted.
In case the resistor is burnt, it must be replaced according to the 
circuit diagram and parts list (in Section 3.8), and further, the 
sensitivity of the tester must be adjusted in the following manner.

(1) Set the polarity select switch to 

+DC,AC,

Ω

 and the measuring 

range select switch to the “DC 0.3 V” range. 

Apply  “DC  0.3  V”  and  turn  the  semifixed  resistor  VR

1

 (as 

indicated on the print board) so that the pointer may indicate the 
full scale value correctly.

(2) Set the measuring range select switch to the “AC 120 V” range. 

Apply “AC 120 V” and turn the semifixed resistor VR

2

 so that the 

pointer may indicate the full scale value correctly.

Above procedures make a complete adjustment on the DC, AC, and 

Ωcircuits.

5.2 Fuse Replacement

CAUTION

There is a risk of electrical shock to persons.

· Prior to replacing the fuse, disconnect the instrument from the 

circuit to be measured, if connected, to prevent an electrical 
shock hazard.

· Do not operate the instrument when the case is left open.

· Use only the specified fuse.

In case the fuse has blown, remove the screw of the center of 
the bottom case. Then, replace the blown fuse with a new one (of 

specified rating) that is provided on the side surface of the case. 

After Services

When the instrument does not operate properly and needs repair, 
please contact Yokogawa or the store where you bought the 
product.

Summary of Contents for 320110

Page 1: ...User s Manual Model 320110 Circuit Tester IM 3201 E 13th Edition July 2014 KP...

Page 2: ...Thank you for purchasing this Model 320110 Circuit Tester To fully exploit and properly operate the instrument read this manual carefully prior to using the product...

Page 3: ...s instrument for the circuit exceeding 250 V including power lines Such strong electrical circuits may contain a surge voltage which is several times stronger than that of circuit voltage thus creatin...

Page 4: ...2 All rights are reserved by Yokogawa No one is allowed to make copies of reprint any part or all the contents of this instruction manual without permission from Yokogawa 3 If any questions arise or e...

Page 5: ...requency Output Level 8 3 6 Resistance Measurement 8 3 7 Checking Battery Amount and Battery Replacement 9 3 8 Circuit Diagram Parts List 10 4 Use of Accessories Additional Specifications 11 4 1 Model...

Page 6: ...iv IM 3201 E...

Page 7: ...a tester without amplifier It can also be used as an electronic voltmeter 3 Results of the measurement with 320110 are highly accurate with 2 for DC or 3 for AC The scale has a built in mirror 105 mil...

Page 8: ...ointer 3 Meter Cover 6 Polarity Select Switch 12 Case 14 Bottom Case Fixing Screw 13 Rubber Leg 4 Cover Fixing Screw 5 Zero Adjust Screw 7 Zero ohm Adjust Knob 8 Measuring Range Select Switch 9 Common...

Page 9: ...resistance measurement it is turned to DC AC This switch has a 3 contact switch to make deflecting angles of the pointer wider to clarify the polarity selection Center Position When the switch is rest...

Page 10: ...g Screw When replacing batteries remove this screw to open the back cover 15 Measuring Probe This instrument provides two lead wires i e one black and one red The clip adaptor can be used conveniently...

Page 11: ...burned out Checking Method Set the measuring range select switch to range The fuse operates normally if the pointer points at zero when shorting the V A terminal The pointer does not deflect when the...

Page 12: ...ch positioning are the same as in the case of the DC current measurement described above 2 Set the measuring range select switch to a desired position on the DC V range Check polarity and connect the...

Page 13: ...es follow instructions described in the paragraph AC Voltage Measurement 7 To the OUTPUT terminal a 0 2 F capacitor 400 W V is internally provided in series so that the DC component is impeded and onl...

Page 14: ...ing the resistance in the circuit of equipment the power source to the equipment must be disconnected For the measurement of a circuit which includes capacitors the charged electricity must be dischar...

Page 15: ...larity noted on the battery holder case 5 Current for each measuring range is as follows at zero position of resistance scale 1 range approximately 150 mA 100 range approximately 1 5 mA 10 k range app...

Page 16: ...9 M F R13 RG08V3C 18 M F R14 RG08V3C 90 M F R15 RG08V3C 9 M F R16 RG08V3C 1 8 M F R17 RD1 2PX 900 k F R18 RD1 4PX 180 k F R19 RD1 4PX 90 k F R20 RD1 4PX 11 3 k F R21 RD1 4PX 10 k F R22 RD1 4PX 10 k F...

Page 17: ...air of gloves e g rubber gloves to ensure safety Avoid measuring high voltage on the power line to prevent possible hazards 1 This probe is intended to be used in the adjustment of high voltage electr...

Page 18: ...20510 Clip On Current Transformer Order Discontinued Item black red Fig 4 3 Clip On Current Transformer 1 By use of this transformer currents to the commercial frequencies can be measured without open...

Page 19: ...ay indicate the full scale value correctly 2 Set the measuring range select switch to the AC 120 V range Apply AC 120 V and turn the semifixed resistor VR2 so that the pointer may indicate the full sc...

Page 20: ...f full scale value Input Impedance 10 k V The tolerance for the low frequency output has been expressed as a percentage of the value obtained by converting the maximum scale value dB to a voltage The...

Page 21: ...lowable Input Voltage DC 15 kV Model 320410 12A DC Shunt Order Discontinued Item Rating Current DC 12 A Rated Voltage Drop 220 mV Accuracy 3 of rating when connected to the tester 320110 Model 320510...

Page 22: ...16 IM 3201 E MEMO...

Page 23: ......

Page 24: ......

Reviews: