3. Operation

3.1 Instruction for Use

1.  Make sure prior to measurement that the pointer rests at the 

zero position of the scale. If not, turn the zero adjust screw to 
set it to zero. Look at the scale so that you can see the pointer 

is accurately aligned with the pointer reflected on the mirror, and 

then read the value.

2.  When the measuring value is unknown, start measurement with 

the maximum range provided and then continue to lower range 
until proper range is obtained. 

Do not select the switch while 

the pointer is still deflecting.

 

3.  Do not place the tester in strong magnetic fields or on steel 

plates to avoid possible pointing errors.

4.  When storing or carrying the tester, turn off the measuring range 

select switch and set the polarity select switch to its center 
position to avoid any possible damage.

Make sure to read and observe all the following cautions before 

starting measurement to ensure safety.

CAUTION

 Checking Fuse

Make sure prior to measurement that the fuse is not burned 
out.

<Checking Method> 

Set the measuring range select switch to 

“Ω”  range. The  fuse  operates  normally 

if  the  pointer  points  at  zero  Ω  when 

shorting the V.A.Ω terminal. 

The pointer does not deflect when the fuse is burned out. 
At the start of measurement, always be sure to check it. 

Never use the instrument when the fuse is burned out 

to avoid electrification. 

It happens when the instrument 

falsely recognizes voltage-applied measured objects as being 
inactive.

 Checking Measurement Range

· Make sure that the measuring range select switch is 

set properly and that the measurement leads and the 
measurement (input) terminal are connected appropriately. 
Measuring high voltages erroneously while the range is 

in  “DC mA”  may  cause  personnel  injury  or  damage  to  the 

instrument.

· Prior to turning on/off the measuring range select switch, 

disconnect the measurement leads from the measured object.

 Installation Location

Be sure to place the instrument on a flat stable surface in a 

stable environment. 
T h e  m e a s u r e m e n t  l e a d s  m a y  g e t  d i s c o n n e c t e d  f r o m 
the measurement terminals due to overturn or fall of the 
instrument. If it happens, there is a possibility of electrical 
shock especially when a high voltage is measured.

Summary of Contents for 320110

Page 1: ...User s Manual Model 320110 Circuit Tester IM 3201 E 13th Edition July 2014 KP...

Page 2: ...Thank you for purchasing this Model 320110 Circuit Tester To fully exploit and properly operate the instrument read this manual carefully prior to using the product...

Page 3: ...s instrument for the circuit exceeding 250 V including power lines Such strong electrical circuits may contain a surge voltage which is several times stronger than that of circuit voltage thus creatin...

Page 4: ...2 All rights are reserved by Yokogawa No one is allowed to make copies of reprint any part or all the contents of this instruction manual without permission from Yokogawa 3 If any questions arise or e...

Page 5: ...requency Output Level 8 3 6 Resistance Measurement 8 3 7 Checking Battery Amount and Battery Replacement 9 3 8 Circuit Diagram Parts List 10 4 Use of Accessories Additional Specifications 11 4 1 Model...

Page 6: ...iv IM 3201 E...

Page 7: ...a tester without amplifier It can also be used as an electronic voltmeter 3 Results of the measurement with 320110 are highly accurate with 2 for DC or 3 for AC The scale has a built in mirror 105 mil...

Page 8: ...ointer 3 Meter Cover 6 Polarity Select Switch 12 Case 14 Bottom Case Fixing Screw 13 Rubber Leg 4 Cover Fixing Screw 5 Zero Adjust Screw 7 Zero ohm Adjust Knob 8 Measuring Range Select Switch 9 Common...

Page 9: ...resistance measurement it is turned to DC AC This switch has a 3 contact switch to make deflecting angles of the pointer wider to clarify the polarity selection Center Position When the switch is rest...

Page 10: ...g Screw When replacing batteries remove this screw to open the back cover 15 Measuring Probe This instrument provides two lead wires i e one black and one red The clip adaptor can be used conveniently...

Page 11: ...burned out Checking Method Set the measuring range select switch to range The fuse operates normally if the pointer points at zero when shorting the V A terminal The pointer does not deflect when the...

Page 12: ...ch positioning are the same as in the case of the DC current measurement described above 2 Set the measuring range select switch to a desired position on the DC V range Check polarity and connect the...

Page 13: ...es follow instructions described in the paragraph AC Voltage Measurement 7 To the OUTPUT terminal a 0 2 F capacitor 400 W V is internally provided in series so that the DC component is impeded and onl...

Page 14: ...ing the resistance in the circuit of equipment the power source to the equipment must be disconnected For the measurement of a circuit which includes capacitors the charged electricity must be dischar...

Page 15: ...larity noted on the battery holder case 5 Current for each measuring range is as follows at zero position of resistance scale 1 range approximately 150 mA 100 range approximately 1 5 mA 10 k range app...

Page 16: ...9 M F R13 RG08V3C 18 M F R14 RG08V3C 90 M F R15 RG08V3C 9 M F R16 RG08V3C 1 8 M F R17 RD1 2PX 900 k F R18 RD1 4PX 180 k F R19 RD1 4PX 90 k F R20 RD1 4PX 11 3 k F R21 RD1 4PX 10 k F R22 RD1 4PX 10 k F...

Page 17: ...air of gloves e g rubber gloves to ensure safety Avoid measuring high voltage on the power line to prevent possible hazards 1 This probe is intended to be used in the adjustment of high voltage electr...

Page 18: ...20510 Clip On Current Transformer Order Discontinued Item black red Fig 4 3 Clip On Current Transformer 1 By use of this transformer currents to the commercial frequencies can be measured without open...

Page 19: ...ay indicate the full scale value correctly 2 Set the measuring range select switch to the AC 120 V range Apply AC 120 V and turn the semifixed resistor VR2 so that the pointer may indicate the full sc...

Page 20: ...f full scale value Input Impedance 10 k V The tolerance for the low frequency output has been expressed as a percentage of the value obtained by converting the maximum scale value dB to a voltage The...

Page 21: ...lowable Input Voltage DC 15 kV Model 320410 12A DC Shunt Order Discontinued Item Rating Current DC 12 A Rated Voltage Drop 220 mV Accuracy 3 of rating when connected to the tester 320110 Model 320510...

Page 22: ...16 IM 3201 E MEMO...

Page 23: ......

Page 24: ......

Reviews: