4.2.5.Calibration description:

(a) This meter can adopt random one-point, two-point or three-point automatic calibration,

after the first point calibration (see item 4.2.2.), press

to exit calibration

mode and enters into measuring mode. The indication icon

will appear on

the lower left corner of LCD. When the measuring accuracy is

≤±

0.1pH, user just

need to choose one kind buffer solution to take one-point calibration according to the

measuring range.

(b) After the second point calibration, (see item 4.2.3.), press

to exit calibration

mode and enter into measuring mode. The indication icon

for

tow-point calibration will appear on the lower left corner of LCD. User can choose

pH4.00 and pH 7.00 to calibrate if the measurement is just within the acidity range

and choose pH 7.00 and pH 10.01 to calibrate if just within the alkalinity range.

(c) User should choose three-point calibration so as to reach a more accurate

measurement if the measuring range is wide, or if the electrode has been used for long

or has ageing phenomenon. As to the new electrode which be used for the first time, it

must be calibrated by three-point calibration to keep the unanimity of the meter slope

adjustment with the pH electrode.

4.3. PH Solution Test:

Immerge pH electrode into the sample solution after washing and dry it, rock the

electrode holder and still, when the LCD appears the icon

to take the reading

after displaying value to be stable.

Note: According to the pH equal temperature measuring theory: the closer the

temperature of the sample solution with the calibration solution, the more accurate
the measuring value will be acquired. So please comply with this theory.

Summary of Contents for P613 PH

Page 1: ...Shanghai Yoke Instrument Co Ltd Model P613 PH Conductivity Meter Operation Manual 1...

Page 2: ...P H Measurement 6 4 1 Preparation Work 6 4 2 Meter Calibration 6 4 3 PH solution Test 8 4 4 Parameter Setting 8 4 5 Considerations 14 4 6The Self diagnose Information 14 5 MV Measurement 14 6 Conduct...

Page 3: ...itable for the trade such as the mining industry power plant water treatment projects and environmental protection etc especially has more extensive application in the field and spot test Built in mic...

Page 4: ...dopts SMT film covering technology to improve meter s production reliability 1 6 Meter has the back light LCD display monitor 1 7 IP54 dust and water resistant 2 Technical Parameters 2 1 pH Measuring...

Page 5: ...mpensation range 0 100 manual auto Electrode constant 0 1 1 10 cm 1 Benchmark temperature 25 20 and 18 2 4 Other Technical Parameters Data storage 150 groups Storage content Series number measuring va...

Page 6: ...icon RM reading to be recalled icon Measuring unit Temperature measuring value and unit Measuring value to be stable icon Electrode calibration indicate icon Indication icon of low voltage appears whe...

Page 7: ...nd data deleting key a When in the measuring state short press less than 1 5s to open or close the back light display b When the storage records are queried long press more than 5s to delete the saved...

Page 8: ...t corner of the LCD and the complete measuring information Again press or key meter will in turn recall all the measuring information b In the recalling mode press key to return to the measuring mode...

Page 9: ...and dry it then immerge it into the pH4 00 buffer solution rock the electrode holder and then still waiting for a stable data and appear then again press key the LCD will appear a twinkling 4 00pH mak...

Page 10: ...st within the acidity range and choose pH 7 00 and pH 10 01 to calibrate if just within the alkalinity range c User should choose three point calibration so as to reach a more accurate measurement if...

Page 11: ...kly Press key determines the current settings and returns Prompt Mark Parameter Setting Items Parameters P1 Manual temperature compensation setting 0 99 9 P2 Resolution selection 0 1pH 0 01pH P3 pH st...

Page 12: ...ndard buffer solution selection P3 a press key in mode P2 to choose mode P3 press key enters into mode P3 see picture 4 4 b Press key or change buffer solution selection there are Europe U S A series...

Page 13: ...4 4 6 Purified water Ammonia added purified water pH restitution P5 a Press key in mode P4 to choose mode P5 press key enters into P5 see picture 4 6 b Press key and key modified OFF no compensation H...

Page 14: ...r off time press key determines and returns 0 means to shut off auto power function c Press key to enter into next parameter setting or press key return to measuring mode 4 4 9 Restore to producer set...

Page 15: ...out the electrode and wash it in purified water before measuring Insert the electrode and screw tight the capsule after measuring to prevent the solution leaking If the soaking solution is turbid or...

Page 16: ...l decrease and its response is getting slow the reading are not correct It could adopt the following method for various cases a The glass bulb is contaminated and aging Put the electrode into 0 1mol L...

Page 17: ...stable icon 2 00 pH or the 19 99 pH this icon appeared when the value has surpassed the measuring range There will also appear such signs when the electrode is not well connected with the meter or wh...

Page 18: ...o be used normally so we should replace the new electrode 6 1 3 Connect the DJS 1 conductivity electrode and temperature electrode into the meter 6 2 Meter s Calibration Standard solution calibration...

Page 19: ...is 10 5 select the 10 electrode type first then set the constant 1 05 that is 10 5 10 1 05 6 3 Conductivity solution Test 6 3 1 Wash the conductivity electrode and throw off the water on it then inser...

Page 20: ...12 88 mS cm 111 9 mS cm Calibration indication icon 6 4 3 There are two kinds of method which are standard solution calibration method and constant calibration method for conductivity electrode calibr...

Page 21: ...in the parameters setting of P5 At the same time meter will do automatic non linear temperature compensation in the high purified water which less than 10 S cm Note When the coefficient of temperatur...

Page 22: ...Prompt Mark Parameter Setting Items Parameter P1 Electrode constant selection 0 20 5 00 P2 Setting of constant coefficient of electrode 0 1 1 0 10 P3 Manual temperature compensation setting 0 99 9 P4...

Page 23: ...b Press or key to change temperature value long press these two settings can be changed quickly press key determine the current set value c Press key to enter into the next parameter setting or press...

Page 24: ...to determine the current selection c Press key to enter into next parameter setting or press key to confirm and return to measuring mode 6 5 8 Selection of temperature units P7 a Under P6 mode to pre...

Page 25: ...d return to measuring mode 6 5 11 Restore to producer setting P10 a Under P9 mode to short press key choose P10 press enter into P10 mode shown as picture 6 10 b Press or key to choose then press key...

Page 26: ...platinum black only can wash it in water by shaking in case to damage the platinum black coating for organic stained things can be washed with hot water with detergent or alcohol 6 6 5 Conductivity e...

Page 27: ...ain replace the parts or products under normal using circumstances from purchased time within one year caused by manufacturing bad and unable to work 8 2 Attached electrodes do not belong to this warr...

Reviews: