background image

9

Figure 3-3. 3rd System Analyzer Display

Vacuum Mode

Note: If you are using a new vacuum sensor for 

the first time, the vacuum sensor calibration num-

ber must be entered. The instrument will report 

incorrect data if this is not done. Go to ‘Vacuum 

Sensor Calibration Utility’ further down in this 

chapter for instructions.

The instrument can simultaneously use display 

inputs from two sources:

Vacuum Sensor

Internal Pressure Transducers

Note: Although the internal pressure transducers 

are primarily used for positive pressures, they also 

give a helpful indication of vacuum activity before 

the vacuum sensor indicates values in the ranges 

that are typical when operating vacuum pumps.

This screen can be accessed from the main menu, 

“Vacuum Gauge” selection, or the “Digital Manifold” 

screen by simply plugging in a vacuum sensor if 

the “Auto   micron meter” is set to “ON”. (See “Set-

tings - Auto Vacuum Gauge” in the Setup menu.)

Figure 3-4. Vacuum Gauge Display

1 Pressure units. While using any of the 

digital manifold or vacuum gauge display 

screens, press the UP/LEFT key to activate 

a pop-up screen where options can be 

changed using the UP/LEFT and DOWN/

RIGHT keys. Press the MENU key to save 

the selections and exit the pop-up screen.

2 Discharge pressure (‘High Side’).
3 When data recording is activated, this 

indicator will gradually change from red 

to blue indicating the size of one page of 

data. During playback, it will help the user 

to decide whether to move through the 

data slowly or page-by-page. Adjusting the 

sampling rate will affect the amount of data 

on each page.

4 Vacuum reading. The message “Out of 

range” will appear until the vacuum reaches 

25,000 microns.

5 Elapsed time display. Press “Clear” to reset 

counter to 0:00:00.

6 Vacuum Units. Vacuum readings can be 

displayed in Microns, mTorr, Torr, mmHg, 

mBar, KPa, and Pa. 

If it is desirable to match the pressure units 

with the vacuum units, the pressure units 

must be selected before proceeding with 

vacuum procedure.

While using any of the digital manifold or 

vacuum gauge display screens, press the 

UP/LEFT key to activate a pop-up screen 

where options can be changed using the 

UP/LEFT and DOWN/RIGHT keys. Press 

the MENU key to save the selections and 

exit the pop-up screen.

7 When recording is activated, the sampling 

rate is displayed here. See “Chapter 4 – 

Data Logging” to set the sampling rate.

8 The minimum and maximum pressures 

encountered since the memory was last 

cleared (“Clear” key). (Typical for high and 

low sides.)

9 Left Side: Suction pressure (‘Low Side’). 

Press the UP/LEFT key to activate a pop-

up screen where options can be changed 

using the UP/LEFT and DOWN/RIGHT 

keys. Press the MENU key to save the 

selections and exit the pop-up screen.

Summary of Contents for 40812

Page 1: ...ached to the instrument YELLOW JACKET Refrigeration System Analyzer Users Manual UPC 40812 40813 and 40815 Versions 1 06 and higher 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1...

Page 2: ...ions 15 Safety 15 Chapter Title Pg Features and Specifications 15 Instrumentation Specifications 16 Warranty Policy 16 Table of Contents List of Tables Table Title Pg 1 1 Safety Information 3 1 2 Symb...

Page 3: ...ced only by Ritchie Engineering Co or authorized service centers Do not use the instrument if it operates ab normally Protection may be impaired When in doubt have the instrument serviced Refer to war...

Page 4: ...page by page scrolling mode during data log playback Battery Battery connector orientation Chapter 2 Getting Acquainted Introduction This instrument will clearly and accurately report critical inform...

Page 5: ...is detected Once the data is stored unit will turn off Maximizing Battery Life Battery life decays fastest when the DIGITAL MANIFOLD display is selected the vacuum sen sor is attached and the backlig...

Page 6: ...Chart Display This display shows the P T Pressure Tempera ture chart for the refrigerant that is selected in the Set Up display Use the or keys to scroll through the chart Press the Menu key to exit...

Page 7: ...Transducer internal High Side Pressure Transducer internal Temperature Probes The system analyzer will display readings from any K type thermocouple probe with a miniature style connector see picture...

Page 8: ...erheat measurement the T1 probe should be mea suring the suction line temperature Temp 2 T2 The temperature indicated by the T2 thermocouple For proper subcool ing measurement the T2 probe should be m...

Page 9: ...High Side 3 When data recording is activated this indicator will gradually change from red to blue indicating the size of one page of data During playback it will help the user to decide whether to mo...

Page 10: ...for signs of oil before connecting the 69086 vacuum gauge sensor Keep the sensor vertical when possible Connect the gauge directly to the system away from the pump Isolate the pump from the system an...

Page 11: ...and reporting at a later time Figure 4 3 Data Log Files Erase File Use the arrow keys to highlight the data file to be erased Press clear to erase Erase All Files Press clear to erase all data files...

Page 12: ...ry default is MM DD YY Time and Date These settings are modified in a slightly different manner Selecting this setting for modification will open a pop up window Each component of the time and date va...

Page 13: ...as dish soap If the unit encounters enough exposure to liquid chemicals that the chemical wicks between the in strument boot and the case immediately remove the boot and clean the case as described ab...

Page 14: ...tem Analyzer Instrument Boot 40823 System Analyzer Battery Door 40824 System Analyzer Bumper Plugs Sensor Boots 40827 USB cable 69101 Vacuum Sensor Cord Assembly 69217 K Type Clamp Probe 10 ft Softwar...

Page 15: ...ls back to past readings Refrigerant Database 98 NIST refrigerant profiles included upgradeable Data Logging Adjustable time intervals from 0 1 to 10 seconds minimum of 24 hours at 1 0 second interval...

Page 16: ...Company Inc is guaranteed to be free from material and workmanship defects for a period of two years All other YELLOW JACKET products included in this package are guaranteed to be free from material a...

Reviews: