yellow jacket 40812 User Manual Download Page 4

Table 1-1. Safety Information

    

Caution

To avoid damage to equipment, follow these 

guidelines:

Do not allow pressures beyond the specifi-

cations listed in this manual.

Be aware that internal pressures can 

change unintentionally when equipment is 

stored with pressure in the system during 

temperature changes. If sub-cooled liquid 

refrigerant is trapped in a hose or manifold

with no room for expansion, it may result in 

dramatic pressure variations with seemingly 

small temperature changes. Pressures can 

reach high enough levels to cause dam-

age to the instrument’s internal pressure 

transducers. Release liquid refrigerant from 

the hoses and manifold when disconnecting 

from a system.

Refer to cautions supplied with batteries 

acquired for use in this instrument.

Do not attempt to introduce liquid or 

samples heavily laden with oil into the 

instrument.

Read and observe instructions and specifi-

cations related to the batteries used in this 

instrument that may cause damage to it.

Do not use this instrument on systems 

containing leak sealing chemicals. These 

leak sealants can collect and harden in the 

instrument, causing permanent damage.

Table 1-2  Symbols

Important information

Power On/Off

 T1

Temperature 1

 T2

Temperature 2

 REC 

Indicates that the instrument is 

recording readings (data logging)

||><||

Indicates page-by-page scrolling 

mode (during data log playback)

Battery

Battery connector orientation

Chapter 2 

Getting Acquainted

Introduction

This instrument will clearly and accurately report 

critical information needed to properly service re-

frigeration and air-conditioning equipment. With its 

many features, time can be saved and the quality 

of service can be verified through data reports for 

customer satisfaction. Some of the most signifi-

cant features include:

High accuracy and resolution

Very fast and sensitive leak detection

Robust temperature compensation with fault  

detection

Data logging and downloading for reporting and 

analysis

High durability and weather-resistance

Note that these instructions do not cover the 

manifold attached to the instrument.

Turning the Instrument On 

and Off 

Press and release the       key to turn unit on. After 

the logo appears briefly, the instrument will pres-

ent the main menu.

Press and hold the       key (~two seconds) to turn 

unit off.

Figure 2-1. Menu Display

Battery Considerations

The instrument uses eight AA batteries. The 

user may select batteries of the following types: 

Alkaline, AA-Lithium, Ni-MH, Ni-Cd, Li-Pol, Li-Ion 

from the Setup menu (see Chapter 5). Do not mix 

battery types, including rating (i.e., do not mix four 

Ni-MH rated at 1600mAh with four Ni-MH rated 

at 1800mAh). Also, each battery in a set of eight 

must be at the same power state, preferably fully 

charged. The battery life indicator is only accurate 

when the correct battery type is selected in the 

SET-UP menu.

4

Summary of Contents for 40812

Page 1: ...ached to the instrument YELLOW JACKET Refrigeration System Analyzer Users Manual UPC 40812 40813 and 40815 Versions 1 06 and higher 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1...

Page 2: ...ions 15 Safety 15 Chapter Title Pg Features and Specifications 15 Instrumentation Specifications 16 Warranty Policy 16 Table of Contents List of Tables Table Title Pg 1 1 Safety Information 3 1 2 Symb...

Page 3: ...ced only by Ritchie Engineering Co or authorized service centers Do not use the instrument if it operates ab normally Protection may be impaired When in doubt have the instrument serviced Refer to war...

Page 4: ...page by page scrolling mode during data log playback Battery Battery connector orientation Chapter 2 Getting Acquainted Introduction This instrument will clearly and accurately report critical inform...

Page 5: ...is detected Once the data is stored unit will turn off Maximizing Battery Life Battery life decays fastest when the DIGITAL MANIFOLD display is selected the vacuum sen sor is attached and the backlig...

Page 6: ...Chart Display This display shows the P T Pressure Tempera ture chart for the refrigerant that is selected in the Set Up display Use the or keys to scroll through the chart Press the Menu key to exit...

Page 7: ...Transducer internal High Side Pressure Transducer internal Temperature Probes The system analyzer will display readings from any K type thermocouple probe with a miniature style connector see picture...

Page 8: ...erheat measurement the T1 probe should be mea suring the suction line temperature Temp 2 T2 The temperature indicated by the T2 thermocouple For proper subcool ing measurement the T2 probe should be m...

Page 9: ...High Side 3 When data recording is activated this indicator will gradually change from red to blue indicating the size of one page of data During playback it will help the user to decide whether to mo...

Page 10: ...for signs of oil before connecting the 69086 vacuum gauge sensor Keep the sensor vertical when possible Connect the gauge directly to the system away from the pump Isolate the pump from the system an...

Page 11: ...and reporting at a later time Figure 4 3 Data Log Files Erase File Use the arrow keys to highlight the data file to be erased Press clear to erase Erase All Files Press clear to erase all data files...

Page 12: ...ry default is MM DD YY Time and Date These settings are modified in a slightly different manner Selecting this setting for modification will open a pop up window Each component of the time and date va...

Page 13: ...as dish soap If the unit encounters enough exposure to liquid chemicals that the chemical wicks between the in strument boot and the case immediately remove the boot and clean the case as described ab...

Page 14: ...tem Analyzer Instrument Boot 40823 System Analyzer Battery Door 40824 System Analyzer Bumper Plugs Sensor Boots 40827 USB cable 69101 Vacuum Sensor Cord Assembly 69217 K Type Clamp Probe 10 ft Softwar...

Page 15: ...ls back to past readings Refrigerant Database 98 NIST refrigerant profiles included upgradeable Data Logging Adjustable time intervals from 0 1 to 10 seconds minimum of 24 hours at 1 0 second interval...

Page 16: ...Company Inc is guaranteed to be free from material and workmanship defects for a period of two years All other YELLOW JACKET products included in this package are guaranteed to be free from material a...

Reviews: