CABLE AND CONTACTS SPECIFICATIONS
:
Pin No.
Colors
Functions & Specs
Power/Level cable(5 cores,
AWG18 /105
℃
/ 220VAC/1.0A,length 2.0 meters
)
A1.
BLUE
External power line
A2.
WHITE
External power line
A3.
GREEN
External power line ground (No connect to power source)
A4.
RED
Level sensor/switch (No connect to power source)
A5.
RED/WHITE
Level sensor/switch (No connect to power source)
I/O cable (13cores,
AWG26/105
℃
/60V DC/0.5A,length 2.0 meters
)
B1.
RED
External voltage signal 0~10V
(Input resistance approx. 20KΩ)
B2.
ORANGE
External current signal 4~20mA
(Input resistance approx. 200Ω)
B3.
YELLOW
External pulse signal
1.
voltage range =0~60V
2.
No-voltage contact or open collector, input resistance
approx. 2kΩ
3.
Max. number of pulses= 9000 pulse/min, min. pulse
width: 3.33 mSec (ON/OFF time) (Note1)
4.
Volt< 0.8V = Low level
,
Volt= 3.0~ 60.0V = Hi level
B4.
BLACK
common
(for all signal of voltage/current/pulse/on-off)
B5.
GREEN/
WHITE
Power supply
Output 5.0 V, max.400mA (with 0.5A auto reset fuse)
B6.
GREEN
External RUN/STOP switch connector
B7.
BLUE
External
“
Full speed
”
switch connector
(The same with press ▲ ▼ on panel simultaneously)
B8.
PURPLE
Pulse signal (dry connection, Note3)
1.
Simultaneous with pulse
2.
Max voltage 60V,(D.C.only)
3.
Continuous Current @ TC=25
℃
= 12A.
4.
Power Dissipation @TC=25
℃
= 20W
Pulse
signal
output
(Pin8,9)
B9.
PURPLE/
WHITE
Pulse signal common
B10.
GRAY
Batch signal (dry connection, Note3)
1.
Max voltage 60V,(D.C.only)
2.
Continuous Current @ TC=25
℃
= 12A.
3.
Power limit @TC=25
℃
= 20W
Batch
Signal
output
(Pin10,11)
B11.
GRAY/WHITE Batch signal common
B12.
BROWN
Low level alarm (dry connection, Note3)
1.
Max voltage 60V,(D.C.only)
2.
Continuous Current @ TC=25
℃
= 12A.
3.
Power limit @TC=25
℃
= 20W
Low
Level
alarm
output
(Pin12,13)
B13.
BROWN/
WHITE
Low level alarm signal common