YCO-1126-01-08
PPC-300 & PPC-300D Technical Manual
IV Setup
16
Parameter Tables
Factory Parameters
Parameter
Function
Default
Settings
88
Span adjustment time delay
002
000: No delay
001 - 005: Delay of specified seconds
90
Mechanical zero 1
N/A
000 - 255: Automatically set during cali-
bration, do not change
91
Mechanical zero 2
N/A
000 - 255: Automatically set during cali-
bration, do not change
92
Mechanical zero 3
N/A
000 - 255: Automatically set during cali-
bration, do not change
93
Span coefficient 1, small
N/A
000 - 255: Automatically set during cali-
bration, do not change
94
Span coefficient 2, small
N/A
000 - 255: Automatically set during cali-
bration, do not change
95
Span coefficient 3, small
N/A
000 - 255: Automatically set during cali-
bration, do not change
96
Span coefficient 1, large
N/A
000 - 255: Automatically set during cali-
bration, do not change
97
Span coefficient 2, large
N/A
000 - 255: Automatically set during cali-
bration, do not change
98
Span coefficient 3, large
N/A
000 - 255: Automatically set during cali-
bration, do not change
99
Region number and gravity
for span adjustment
N/A
000 - 150: Automatically set during cali-
bration, do not change
A0
Board sensitivity adjustment
1
N/A
000 - 099: Board sensitivity adjustment
mantissa, x 1000
A1
Board sensitivity adjustment
2
N/A
000 - 255: Automatically set during board
sensitivity adjustment, do not
change
A2
Board sensitivity adjustment
3
N/A
000 - 255: Automatically set during board
sensitivity adjustment, do not
change
A5
Display of "lb:oz" unit
001
000: Enabled
001: Disabled
B0
Factory setting
*
000: Do not use
001 - 005: UDS-1V versions (Japan)
006 - 019: For specific OEM versions
* Value determined by capacity, see capacity specific table after parameter tables.