4. Operation Method
Temperature Output Terminal
Specification
Temperature Output
(ANALOG)
The curent (mA) corresponding to the measured temperature is output.
Output temperature range: -5
~
55
℃
Output voltage: 4
~
20mA
Load impedance: 600Ω or less
Resolution:±1
℃
Connection: M4 screw terminal block
Alarm Output
(
ALARM
)
Output when an abnormality is detected
(
See “Safety device and error
codes “ on P.48 for description of a specific error.
)
a contact
(
Relay contact
)
Contact capacity
:
AC250V
3A
:
DC30V
3A
Connect to
:
M4 screw terminal block
(
common
)
Temperature
(℃)
Output current
(
mA
)
-5
4.00
5
6.67
15
9.33
25
12.00
35
14.67
45
17.33
55
20.00