22
4. Operating procedures
Operating procedures (auto start operation)
MEASURED TEMP.
℃
HEATER
ALARM
AUTO STOP
AUTO START
ENTER
SET
TEMP.
OVER TEMP.
PROTECTOR
FIXED TEMP.
RUN
STOP
SUB
MENU
TIMER
FIXED
TEMP.
PROGRAM
PROGRAM
3. Stopping and ending timer operation
Operation starts automatically when the set time comes.
Press the RUN/STOP key for about one second to start or stop
operation. The screen switches to the initial setting screen.
When you want to
correct set
temperature or set
time, or confirm
settings
When you want change the set temperature or a set time, press the
TIMER key during operation, set temperature and time for the auto
start operation with the
▼ ▲
keys and press the ENTER key to
complete.
Note, however, when you change the set time you need to set a time
calculated by adding the time already passed to the time to be added.
Pressing the
▼
key during operation will display the set temperature,
the operation mode and the remaining time on the SET TEMP. display.
When operation has started after the auto start time, you cannot
change the set time. In this case, first stop operation with the
RUN/STOP key and repeat all settings.