6
2. Before operating the unit
Precautions when installing the unit
Warning
9. Be sure to connect the ground wire.
・
When the unit has no ground terminal, class D grounding work is necessary and please
consult your dealer or our customer service center.
・
Securely connect to a distribution board or an outlet.
We recommend use of a ground type outlet
tap.
When a bipolar type outlet tap is used
アース線
2極タイ プ
コン セン ト
Insert the ground adaptor included as an option,
into a power plug confirming the polarity of the
outlet. Connect the grounding wire (green) of the
ground adaptor to the ground terminal on the
power supply equipment.
When there is no ground terminal.
In this case, class D grounding work (100
Ω
or less) is necessary and please consult your
dealer or our customer service center.
Do not connect the grounding wire to any parts or lines other than a correct grounding
terminal such as a gas pipe, a water pipe or a telephone line.
Otherwise, an accident or a malfunction may result.
10. Use a dedicated outlet for the power supply.
Use a power supply that meets the electrical capacity of the unit.
Electrical capacity:
:
AC115V
50/60Hz
5.1A
*
When the unit will not start even when you turn the refrigerator switch
to “ON”, check
for abnormal main voltage or if the unit is connected to the same power supply line as
other devices and connect it to another line if necessary. Avoid using wire extensions
using a cord reel, which may result in degradation of freezing capacity or stoppage of
refrigerator from possible voltage drops. Do not connect too many wires to a single
outlet using a branching adaptor. Otherwise, a fire or an electric shock may result.
Ground wire
Bipolar
receptacle
G
G
Ground type tap
Power plug
Summary of Contents for CA301
Page 2: ......