Reference) About the Service Mode Menu

The Service Mode has six service menu items: OP1 to
OP6.    However, only menu item OP2 is used in this
calibration.
OP2:  Calibration menu

This operation involves calibrating the zero level of
the CO

2

 concentration transmitter.    When this

operation is executed, calibration is performed so that
the CO

2

 concentration measured on the transmitter at

this time is taken as zero.    When you execute this
operation, be sure to follow the above calibration
method.

If you hold down the (-) button in the service mode for about
three seconds, that service mode menu item (see reference
page) is executed.    (All LEDs (green, yellow, red) light.)
Do not execute menu items other than OP2.

Explanation of Pushbuttons

Button

Description

(

) Button

Used to enter the service mode, and to advance the menu item in the service mode.

(

) Button

Used to enter the service mode, and execute the operation in the service mode.

Explanation of LEDs

Standard mode

Service mode

Green LED

Blinks during normal operation

Same as standard mode

Yellow LED

Lights when an error occurs.

Lights when the service mode is entered.

Red LED

Lights when the output range is exceeded.

Lights during execution of a service mode
menu item.

CO

2

 concentration zero gas inlet

(

) button

(

) button

Yellow LED

Red LED

Green LED

LCD panel