12
3.2 Connecting air piping
WARNING
- Before starting work, make sure that the air compressor is shut off.
1) Connect an air valve, air filter, regulator and if necessary lubricator (hereinafter called the
"peripheral equipment") to hose which connected to compressor. Refer
<
NOTE
>
for detail
information.
2) Install these peripheral items supported by brackets, etc., near the pump.
3) Connect the hose from the peripheral equipment to the air valve of the pump's supply port.
CAUTION
- Use a flexible hose to absorb pump vibration, and ground the hose.
- Make sure that there will be no external force on any connection part of the pump. Be
especially careful not to have the pump support part of the weight of the hose and the
piping.
- The piping and the peripheral equipment may become clogged with sludge. Clean the inside
of the piping for 10 to 20 seconds before connecting it to the pump.
- Be sure to sufficiently ground the piping and peripheral equipment.
- For air purging, use clean air such as N
2
because the purge air will come in contact with the
powder material being pumped.
<
NOTE
>
- Air piping size should be equal to the inlet port of the pump to supply enough air to run the pump.
The air compressor should be able to provide adequate air flow to the pump. Attach the compressor
as close as possible to the pump, considering operability and stability of air pressure.
- If you use a solenoid valve as the air valve, be sure it is a three-way valve. When the valve is closed,
the internal compressed air of the pump will be released, and this will switch the spool to its normal
position.
- Use of a coupler for the connection part of each hose will make operation and maintenance easier.
- If you use the pump intermittently the pump will not require lubrication. However lubrication is
recommended if running the pump continuously for long periods or using very dry air or at high
temperatures. This will guarantee the life of the pumps seals.
*Continuous operation: When the pump operates continuously for longer than 1 hour and is stopped
for less than 15 minutes.
*Lubrication: Use only turbine oil Class 1 grade oil (equivalent to ISO VG 32), under the following
conditions; Oil concentration at 50mg/m³, Absolute pressure at 0.1MPa. Maximum temperature of
20°C and Humidity at 65%.
Fig3.2