![YaliTech LAQUAact-PC110 Instruction Manual Download Page 24](http://html.mh-extra.com/html/yalitech/laquaact-pc110/laquaact-pc110_instruction-manual_894427024.webp)
Basic operation
8
• Setting mode
<Channel 1>
Screen
Layer
Description
P1: PH
pH measurement settings
P11: BUF
Selection of standard solution
type for calibration:
USA, NIST, CUST
P12: CAL DATA
Calibration data display
P13: CAL ALR
Calibration alarm setting:
1 day to 400 days
P14: CAL CLR
Deletion of calibration data
P2: TEMP
Temperature setting
Selection of temperature
conversion or display
: ATC, MTC
P3: GEN
General settings
P31: MEAS
Selection of auto hold type:
auto stability, auto hold
P32: AUTO OFF
Automatic power off setting:
0 min to 30 min
P33: RESET
Initialization of settings
P34: DATE
Date and time setting
P35: PRINT
Test print