Other settings
91
< Calibration data (pH, inspection was performed) >
< Calibration data (pH, inspection was not performed) >
Printout format
Description
Inst. model
:
LAQUAact-PC110 Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
Elect. status
:
GOOD
Electrode status based on calibration result
Offset
:
0.7 mV
Asymmetry potential of calibration data
Sensitivity
Sensitivity of calibration data
pH 4.01 - 6.86
:
98.9%
Repeatability
:
0.01
Repeatability of calibration data
Calibration data
Date
:
2016/01/01
Calibration date
Time
:
09
:
00
Calibration time
pH 4.01
:
176.9 mV
Measurement potential
:
25.0°C ATC
Temperature value and temperature setting
pH 6.86
:
7.6 mV
Measurement potential
:
25.0°C ATC
Temperature value and temperature setting
Inspection before use
pH 6.85
:
7.8 mV
Measurement potential at time of
repeatability inspection
:
25.0°C ATC
Temperature value and temperature setting
Printout format
Description
Inst. model
:
LAQUAact-PC110 Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
Elect. status
:
GOOD
Electrode status based on calibration result
Offset
:
0.7 mV
Asymmetry potential of calibration data
Sensitivity
Sensitivity of calibration data
pH 4.01 - 6.86
:
98.9%
Calibration data
Date
:
2016/01/01
Calibration date
Time
:
09
:
00
Calibration time
pH 4.01
:
176.9 mV
Measurement potential
:
25.0°C ATC
Temperature value and temperature setting
pH 6.86
:
7.6 mV
Measurement potential
:
25.0°C ATC
Temperature value and temperature setting