SensoLyt
®
700 IQ H
Commissioning
11
ba76206e02
12/2020
them dry using compressed air).
NOTE
The pH/ORP sensor SensoLyt
®
700 IQ H may only be submersed in
conjunction with a mounted combination electrode. When changing the
electrode, avoid the penetration of moisture into the pH/ORP sensor,
as the sensor could otherwise be destroyed. Which electrodes can be
used together with the pH/ORP sensor SensoLyt
®
700 IQ H is given in
.
Are the plug
connections dry?
Before connecting the sensor and sensor connection cable, please
make sure the plug connections are dry. If moisture gets into the plug
connections, first dry the plug connections (dab them dry or blow them
dry using compressed air).
Connecting the sensor
to the sensor
connection cable
Do not suspend the sensor on the sensor connection cable.
Use an armature or electrode holder.
Do not suspend the sensor on the sensor connection cable.
Use a sensor holder or armature. Information on this and
other IQ
S
ENSOR
N
ET
accessories is given in the
YSI
cata-
log and on the Internet.
1
Take the protective caps off the plug connections of the sensor
and the SACIQ sensor connection cable and keep them safe.
2
Plug the socket of the SACIQ sensor connection cable onto the
plug head connector of the sensor. At the same time, rotate the
socket so that the pin in the plug head connector (1) clicks into
one of the two holes in the jack.
3
Then, screw the coupling ring (2) of the sensor connection
cable onto the sensor up to the stop.