Commissioning
SensoLyt
®
700 IQ H
10
ba76206e02
12/2020
3
Commissioning
3.1
Scope of delivery
SensoLyt
®
700 IQ H
The sensor is equipped with a protective hood
and protective caps
Operating manual.
3.2
Installation
Safety guidelines for
installation in a
hazardous location
Connection cable
A sensor connection cable of the SACIQ type is required to connect the
sensor. The cable is available in different lengths.
Are the plug
connections dry?
Before connecting the sensor and sensor connection cable, please
make sure that the plug connections are dry. If moisture gets into the
plug connections, first dry the plug connections (dab them dry or blow
GEFAHR
Explosion hazard.
Only hazardous location rated sensor models
("Hazloc Sensor") must be used in hazardous
locations. Verify the rating on the product name
plate on the sensor shaft.
The associated apparatus MIQ/BB1 (YSI part no.
207001Y) has to be used for connection. Refer to the
following documents for proper installation:
Control drawing, YSI document no. 28123
(see MIQ/BB1 operating manual)
MIQ/BB1 operating manual
All applicable electrical code regulations.
GEFAHR
Explosion hazard.
For hazardous location use only the cable types list-
ed in the control drawing must be used.
How to connect the SACIQ sensor connection cable to the
terminal strip of a MIQ module is described in chapter 3
I
NSTALLATION
of the IQ
S
ENSOR
N
ET
system operating man-
ual.