Alyza IQ NH4
Technical data
155
ba76202d02
05/2020
7
Technical data
7.1
Measuring characteristics
NH4-N / NH3-N
,
NH4
Measuring method
Berthelot method in conjunction with an LED photometer
Measuring ranges
and resolution
All specifications concerning the measuring accuracy refer to the use of suitable
standard solutions.
Response time t90
< 10 min
Process variation
coefficient
Detection limit
Measur-
ing
range
Measuring
mode (cita-
tion form)
Measuring
ranges
Reso-
lution
Accuracy
Measurin
g range 1
NH4-N
and
NH3-N
0,02 .. 5,00 mg/L
0,02 .. 5,00 ppm
0.01
±2 %, ±0.02
±2 %, ±0.02
NH4
0,03 .. 6,00 mg/L
0,03 .. 6,00 ppm
0.01
±2 %, ±0.03
±2 %, ±0.03
Measurin
g range 2
NH4-N
and
NH3-N
0,10 .. 20,00 mg/L
0,10 .. 20,00 ppm
0.01
±3 %, ±0.10
±3 %, ±0.10
NH4
0,13 .. 25,00 mg/L
0,13 .. 25,00 ppm
0.01
±3 %, ±0.13
±3 %, ±0.13
Measuring
range
Process variation coefficient
1
< ±2%
2
< ±3%
Measuring
range
Detection limit
1
≤
0.01 mg/L
NH4-N / NH3-N
2
≤
0.03 mg/L
NH4-N / NH3-N