What to do if...
pH/ION 3310
66
ba77052d01
03/2014
No stable measured
value
Obviously incorrect
measured values
Cause
Remedy
Electrode:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Temper if necessary
Ele test sample:
– Conductivity too low
– Use suitable combination elec-
trode
– Temperature too high
– Use suitable combination elec-
trode
– Organic liquids
– Use suitable combination elec-
trode
Cause
Remedy
Electrode:
– Electrode unsuitable
– Use suitable combination elec-
trode
– Temperature difference between
buffer and test sample too great
– Adjust temperature of buffer or
sample solutions
– Measurement procedure not suit-
able
– Follow special procedure