Settings
pH/ION 3310
50
ba77052d01
03/2014
8.1.5 Settings for ISE measurements
The settings for ISE measurements are made in the menu for calibration and
measurement settings of the ISE measurement.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<ENTER>
(or
<F1>
/
[Menu]
)
key. After com-
pleting the settings, switch to the measured value display with
<M>
.
In the following table, only those settings are listed that influence the ISE mea-
surement:
Menu item
Possible setting
Description
Calibration
/
Cali-
bration record
-
Displays the calibration
record of the last calibra-
tion.
Calibration
/
Cali-
bration data stor-
age
/
Display
-
Displays the calibration
record.
Further options:
Scroll through the cali-
bration records with
<
><
>
.
Output the displayed
calibration record to the
interface with
<F2>
/
[USB output]
.
Output all calibration
records to the interface
with
<F2__>
[USB out-
put]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the
measured value display
with
<M>
.
Calibration
/
Cali-
bration data stor-
age
/
Output to USB
-
Outputs the calibration
records to the interface.
Man. temperature
-25 ...
+25
... +130 °C
Entry of the manually
determined temperature.
For measurements without
temperature sensor only.
ISE setup
/
AutoRead criterion
low
medium
high
Selection of the AutoRead
criteria (see section 7.1.1
M