Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the
<F2>
[USB output]
key.
Displaying the
cali-
bration data storage
The calibration records of the last calibrations (up to 10) are available in the
menu,
<ENTER>
/
Calibration
/
Calibration data storage
and in the menu,
<ENTER>
/
Storage & config
/
Data storage
/
Calibration data storage
.
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
or
Data storage
/
Calibration data stor-
age
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
or
Data storage
/
Calibration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.
Display
Calibration
record
Cell constant [cm
-1
]
+++
within the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Error
Error
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
Error elimination (see section 11
W