Multi 3630/3620 IDS
Operation with OxiTop
®
-IDS measuring heads
ba77045e01
09/2019
25
Starting the
measurement
Completing the
measurement
18
If necessary, create another sample in the set with
<F2>
/[
New
].
As the name for the
Sample
, a new number is automatically selected.
The name can be changed freely (max. 8 characters [numerals / let-
ters])
Assign the OxiTop
®
-IDS
measuring heads
to the new sample (repeat
steps 13-17).
19
Using
<ESC>
, or with the menu item
continue
, switch to the
Set
menu
.
20
Using
Start measurement
, start measurement for all samples in the set
at the same time.
The LEDs at the measuring head flash blue when the signal LED on
the Measuring head is switched on.
AutoTemp
is displayed on the display when the
AutoTemp
function is
switched on at the measuring heads.
The settings cannot be changed for a measurement once it is
started!
Each OxiTop
®
-IDS measuring head connected to the meter can
indicate the following data on the display :
name of the set
name of the sample
settings
For this press the <ENTER> key at the measuring head.
21
Keep the sample bottles (with the OxiTop
®
-IDS measuring heads
screwed on) in a place that is controlled to 20°C (e.g. thermostat cabi-
net/box) for the specified number of days.
When the measuring temperature has been reached (function
Auto-
Temp
:
after at least 1 hour, at most approx. 3 hours), the OxiTop
®
-IDS
measuring head automatically starts measuring the oxygen consump-
tion.
22
During the specified measuring duration, stir the sample continuously
(stirring platform, magnetic stirring rods) and keep the temperature
controlled.
The OxiTop
®
-IDS measuring head automatically stores the specified
number of measured values each day during the selected measuring
duration.
23
At the end of a measurement, display the details of the
Set
.
The measurement data of all measuring heads of the set are called up.
24
Using
<M>
,
switch to the
OxiTop set management
.