
Operation
Multi 3630/3620 IDS
16
ba77045e01
09/2019
4.3
Settings
4.3.1 Measurement settings of the set
The measurement settings for a set are defined while the set is being estab-
lished.
Default settings are printed in bold.
Menu item
Possible setting
Explanation
Measuring mode
BOD standard
BOD special
Pressure p
Depending on the selected measuring
mode, some of the following settings are
enabled or grayed out
Set name
max. 8 characters
The 8 characters can be freely selected
Preselected is a clear combination of
numerals (date).
An index is also assigned autimatically.
Duration
30 min,1 h ...1 d...
5
d...
180d
Duration of the measurement
Settings
General settings for the set
AutoTemp
on
off
Function to control the temperature
adjustment. Measurement is started
automatically (at least 1 hour, at most 3
hours after the OxiTop
®
-IDS measuring
head is started).
Pre-tempering the samples to exactly
20°C is not required.
The samples can be inserted with a tem-
perature in the range of 15°C ... 21°C.
The
AutoTemp
function takes over the
temperature control.
Warning difference
pressure
(Only for
Measuring
mode
:
Pressure p
)
50 ...
150
... 500 hPa
Difference pressure (negative or positive
pressure difference compared to the
start pressure). If it is exceeded, the
meter marks the sample with an excla-
mation mark (
!
).
The warning when the specified
Warning
difference pressure
is exceeded is trig-
gered when:
– the measurement data have been
transmitted to the meter and
– the measured value that was last
stored by the OxiTop
®
-IDS measur-
ing head exceeds the difference pres-
sure