Operation
Multi 3430
67
ba75865e19
04/2015
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.9.3
Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview
The following settings are possible for D.O. sensors:
Calibration record (display)
Calibration data storage (display)
Calibration interval
FDO Check
Check interval
Salinity correction
Salinity (salinity equivalent)
Settings
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the
<MENU/ENTER>
key. After
completing the settings, switch to the measured value display with
<M>
.
Measured
parameter
Time interval
Stability in the time interval
D.O. concen-
tration
20 seconds
Δ
: Better than 0.03 mg/l
D.O. saturation
20 seconds
Δ
: Better than 0.4 %
D.O. partial
pressure
20 seconds
Δ
: Better than 0.8 mbar
Temperature
15 seconds
Δ
: Better than 0.5 °C
Menu item
Possible
setting
Description
Calibration
/
Cali-
bration record
-
Displays the calibration record of the
last calibration
Calibration
/
Cali-
bration data
storage
-
Displays the last calibration records
(max. 10)
Summary of Contents for WTW Multi 3430
Page 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Page 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Page 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Page 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Page 133: ......