Multi 3430
Operation
78
ba75865e19
04/2015
Note
You can prematurely terminate the
Stability control
function manually
with
<MENU/ENTER>
at any time. If the
Stability control
function is pre-
maturely terminated, the current measurement data are output to the
interface without the AutoRead info.
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
1
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2
Using
<MENU/ENTER>
, activate the
Stability control
function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.
3
Using
<MENU/ENTER>
, start a further measurement with
Sta-
bility control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured
parameter
Time interval
Stability in the time interval
Conductivity
ϰ
10 seconds
Δ
: Better than
1.0% of measured value
Temperature
10 seconds
Δ
: Better than 0.5 °C
Summary of Contents for WTW Multi 3430
Page 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Page 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Page 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Page 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Page 133: ......