Dissolved oxygen
MIQ/IDS
28
ba77277d01
01/2024
6.2
Check and calibration
6.2.1 General information
Factory calibration
The FDO
®
925
is factory calibrated. The measuring characteristics of the sensor
cap remain stable for the specified service life. Thus, calibration is usually not
required.
When does a check
or calibration make
sense?
A check or calibration can be useful in the following special cases:
If the measured values appear to be implausible and it is assumed that the
service life of the sensor cap is over
Routinely within the framework of the company quality assurance
Principle
A check/calibration is carried out in water vapor-saturated air (test and storage
vessel
FDO
®
Check).
With the calibration procedure, the relative slope of the sensor is determined.
The calibration is evaluated based on the relative slope and the intensity
(successful <-> unsuccessful).
6.2.2 Function check
Procedure
Generally, a check on the IQ S
ENSOR
N
ET
is carried out as follows. For system
specific details, please refer to the IQ S
ENSOR
N
ET
system operating manual.
The check/calibration can be carried out on the following devices:
On the
IQ S
ENSOR
N
ET
on an IDS meter
(see operating manual of the meter)
Details of the check or calibration may vary depending on the device
selected.
Prepare the check and storage vessel FDO
®
Check
.
The sponge in the
FDO
®
Check
check and storage vessel must be
moist (not wet).