MIQ/IDS
pH
15
ba77277d01
01/2024
4
pH
The MIQ/IDS module connects IDS sensors to the IQ S
ENSOR
N
ET
.
Suitable sensors: see section 4.5 A
4.1
Measuring
Prerequisites
The IQ S
ENSOR
N
ET
system is in operation
(see IQ S
ENSOR
N
ET
system operating manual).
The sensor is connected to the MIQ/IDS module
4.2
Calibration
4.2.1 General information on calibration
CAUTION
Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective mea
-
sures must be taken (protective clothing, protective goggles,
etc.).
Calibrate the sensor at regular intervals (depending on the applica-
tion).
1. Install the sensor in a flow through armature and provide the sample flow.
or
Submerse the sensor in the sample.
Calibration can be carried out on the following devices:
on the
IQ S
ENSOR
N
ET
on an IDS meter
(see operating manual of the meter)
Calibration details may vary depending on the device selected.