MIQ/CHV PLUS
Installation
15
ba75539d03
03/2018
3.6
Function check
General
instructions
This check is also suitable for the troubleshooting of malfunctions.
If the valve is not controlled via the IQ S
ENSOR
N
ET
system, the valve control line
must be connected.
To check that the valve triggers correctly, proceed as follows:
WARNING
Danger of injury from lines that are under pressure. Before
opening the compressed air line, ensure that the section
of line concerned is free of pressure. The compressed air
glands in the housing may only be opened by a service
technician authorized by WTW.
1
MIQ/CHV PLUS + IQ S
ENSOR
N
ET
:
Put the IQ S
ENSOR
N
ET
system into operation.
MIQ/CHV PLUS + measuring transmitter:
Supply the power supply module with voltage and put the measuring
transmitter into operation.
2
Adjust the cleaning intervals on the measuring system so that the valve
opens.
or
Open the valve in the
Valve function
menu with the
Manual control
func-
tion (see system operating manual).
3
Check whether there is a compressed air stream on the MIQ/
CHV PLUS.