Conditioning, calibration, measurement
Cl 800
Conditioning, calibration, measurement
General
information
When operating the electrode ensure that
the filling opening (1) for the bridge electrolyte is open
the inner junctions (3) are covered with bridge electrolyte
no air bubbles are in the bridge electrolyte
the depth of immersion is within the optimum range:
Before measuring
4
When doing so refill any spilled bridge electrolyte so that the inner
junctions are covered with bridge electrolyte.
5
Rinse the combination electrode with deionized water.
6
Wipe the shaft using a clean paper towel.
Minimum depth of im-
mersion
The ground junction (4) must be covered
Maximum depth of im-
mersion
Approx. 1 cm below the fluid level (2) of the
bridge electrolyte
1
2
3
4
1
Condition every ion sensitive electrode in standard solution before
using it.
Summary of Contents for wtw Cl 800 DIN
Page 2: ...2 ba15355d04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...6 ba15355d04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15355d04 12 2021 Technische Daten Cl 800...
Page 15: ...OPERATING MANUAL ba15355d04 12 2021 Cl 800 DIN CHLORIDE SENSITIVE ELECTRODE OPERATING MANUAL...
Page 16: ...16 ba15355d04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 24: ...24 ba15355d04 12 2021 Technical data Cl 800...
Page 25: ...MODE D EMPLOI ba15355f04 12 2021 Cl 800 DIN LECTRODE S LECTIVE DE CHLORURE MODE D EMPLOI...
Page 26: ...26 ba15355f04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 34: ...34 ba15355f04 12 2021 Caract ristiques techniques Cl 800...
Page 35: ...MANUAL DE INSTRUCCIONES ba15355s04 12 2021 Cl 800 DIN ELECTRODO SELECTIVO AL CLORURO MANUAL DE...
Page 36: ...36 ba15355s04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 44: ...44 ba15355s04 12 2021 Especificaciones t cnicas Cl 800...
Page 45: ......