Cl 800
Sample preparation
ba15355d04
12/2021
19
Sample preparation
Add 2 % ISA/FK solution. This sample conditioning solution creates optimum
conditions for measuring. It provides a constant ionic strength and similar
diffusion potentials at the reference electrode in standard solution and test
sample.
Response times
The response time depends on the concentration range. It is
several seconds at high concentrations,
several minutes near the detection limit.
The measured value is stable if the value does not change by more than
0.1 mV within 30 seconds.
Interferences
Complexes with Bi
3+
, Cd
2+
, Mn
2+
, Pb
2+
, Sn
2+
, Tl
3+
The conditioning time depends on the previous history of
the ion sensitive electrode. To achieve a good responding
behavior, the concentration of the standard solution should
be similar to the concentration expected of the test sample.
During commissioning and after long periods of storage:
condition the electrode overnight.
Connect the electrode(s) to the meter and switch to the
measured value display (see operating manual of the
meter). The ion sensitive electrode is sufficiently condi-
tioned when the measured value no longer changes consid-
erably under otherwise constant conditions (temperature,
composition of the solution).
2
Remove any air bubbles in the bridge electrolyte by slightly knocking
against the shaft.
3
Calibrate according to the operating manual of the meter and the
analysis specification.
If you would like to have more detailed information concerning
sample preparation and measuring procedures, WTW provides
a large number of application reports for various applications.
Summary of Contents for wtw Cl 800 DIN
Page 2: ...2 ba15355d04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...6 ba15355d04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15355d04 12 2021 Technische Daten Cl 800...
Page 15: ...OPERATING MANUAL ba15355d04 12 2021 Cl 800 DIN CHLORIDE SENSITIVE ELECTRODE OPERATING MANUAL...
Page 16: ...16 ba15355d04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 24: ...24 ba15355d04 12 2021 Technical data Cl 800...
Page 25: ...MODE D EMPLOI ba15355f04 12 2021 Cl 800 DIN LECTRODE S LECTIVE DE CHLORURE MODE D EMPLOI...
Page 26: ...26 ba15355f04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 34: ...34 ba15355f04 12 2021 Caract ristiques techniques Cl 800...
Page 35: ...MANUAL DE INSTRUCCIONES ba15355s04 12 2021 Cl 800 DIN ELECTRODO SELECTIVO AL CLORURO MANUAL DE...
Page 36: ...36 ba15355s04 12 2021 Cl 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 44: ...44 ba15355s04 12 2021 Especificaciones t cnicas Cl 800...
Page 45: ......