What to do if ...
Br 800
Calibration line of a bromide sensitive electrode
What to do if ...
Measured value
unstable
Slope too low
1
0.1
10
100
1000
~58
mV
mg/l Br
-
10
10
10
10
mol/l Br
-
10
Meßkettenspannung
Electromotive force
10-fache Konzentrationsänderung
10-fold concentration change
-5
-6
-4
-3
-2
Cause
Remedy
– Inner junctions not
sufficiently wetted with
bridge electrolyte
– Fill up bridge electrolyte until the inner
junctions are covered with bridge
electrolyte
– Inner junctions encrusted
– Leave the bridge electrolyte to react on
the inner junctions for some hours until
the crusts have dissolved.
– Ground junction
contaminated
– Rinse ground junction with bridge
electrolyte
– Cable broken
– Exchange the electrode
Cause
Remedy
– Membrane surface
contaminated
– Polish the membrane surface with
polishing strip S/FK (see M
– Conditioning time too short
– Extend the conditioning time
– Standard solutions too old
– Use new standard solutions
Summary of Contents for wtw Br 800 DIN
Page 2: ...2 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...6 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15351d04 12 2021 Technische Daten Br 800...
Page 15: ...OPERATING MANUAL ba15351d04 12 2021 Br 800 DIN BROMIDE SENSITIVE ELECTRODE OPERATING MANUAL...
Page 16: ...16 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 24: ...24 ba15351d04 12 2021 Technical data Br 800...
Page 25: ...MODE D EMPLOI ba15351d04 12 2021 Br 800 DIN LECTRODE S LECTIVE DE BROMURE MODE D EMPLOI...
Page 26: ...26 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 34: ...34 ba15351d04 12 2021 Caract ristiques techniques Br 800...
Page 36: ...36 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 44: ...44 ba15351d04 12 2021 Especificaciones t cnicas Br 800...
Page 45: ......