Br 800
Storage
ba15351d04
12/2021
21
NOTE
Never pull the connection cable of the electrode.
The cable might be damaged.
Storage
Between two
measurements
Condition the electrode in standard solution.
Overnight
Condition the combination electrode in standard solution with the filling
opening open. To avoid a contamination of the bridge electrolyte with
standard solution, the level of the bridge electrolyte must be clearly above the
level of the conditioning solution.
For the following measurement, exchange the bridge electrolyte. To do so,
press the shaft of the combination electrode against the connection head and
let the bridge electrolyte drain off. Then fill new bridge electrolyte into the
filling opening.
Longer than
overnight
Remove the bridge electrolyte, then rinse the combination electrode with
deionized water and dab it dry with a clean paper towel. Store the combi-
nation electrode in the storage vessel in a dry place. For the following
measurement, prepare the combination electrode as described in section
C
Recommended accessories
Store the reference electrode according to the instructions in its
operating manual.
Description
Model
Order no.
Bridge electrolyte
ELY/BR/503
106 575
Standard solution 10 g/l Br
-
ES/Br
120 120
ISA sample conditioning solution
ISA/FK
140 110
Polishing strips (24 pieces)
S/FK
180130
Summary of Contents for wtw Br 800 DIN
Page 2: ...2 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 6: ...6 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 14: ...14 ba15351d04 12 2021 Technische Daten Br 800...
Page 15: ...OPERATING MANUAL ba15351d04 12 2021 Br 800 DIN BROMIDE SENSITIVE ELECTRODE OPERATING MANUAL...
Page 16: ...16 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 24: ...24 ba15351d04 12 2021 Technical data Br 800...
Page 25: ...MODE D EMPLOI ba15351d04 12 2021 Br 800 DIN LECTRODE S LECTIVE DE BROMURE MODE D EMPLOI...
Page 26: ...26 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 34: ...34 ba15351d04 12 2021 Caract ristiques techniques Br 800...
Page 36: ...36 ba15351d04 12 2021 Br 800 Copyright 2021 Xylem Analytics Germany GmbH...
Page 44: ...44 ba15351d04 12 2021 Especificaciones t cnicas Br 800...
Page 45: ......