pH value
Lab 865
32
ba77145e02
12/2017
Calibration record
(example)
-
- 30 ... <- 25
or
->+25 ... + 30
-62,0 ... <-61,0
or
>-56,0 ... -50,0
Clean the electrode according to the
electrode operating manual
Error
Error
- 30 ... + 30
-62,0 ... -50,0
Error elimination (see section 11 W
Display
Calibration
record
Zero point
[mV]
Slope
[mV/pH]
Lab 865
Ser. no. 11292113
CALIBRATION pH
01.02.2014 15:55
Ser. no. 10501234
Buffer 1
Buffer 2
7.00
Buffer 3
Voltage 1
184.0 mV
Voltage 2
3.0 mV
Voltage 3
-177.0 mV
Temperatur 1
24.0 °C
Temperatur 2
24.0 °C
Temperatur 3
24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++