Guidance and manufacturer’s declaration –
electromagnetic emissions
The ForeSite SS system is intended for use in the electromagnetic
environments specified below. ForeSite SS system users should ensure
that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic
Environment Guidance
RF emissions CISPR 11
Group 1
The ForeSite SS system
uses RF energy only for
its internal function.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference
in the nearby electronic
equipment.
RF emissions CISPR 11
Class A
The ForeSite SS system
is suitable for use in all
establishments other
than domestic and those
directly connected to the
public lowvoltage power
supply network that
supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Recommended separation distances between
portable and mobile RF communications equipment and the ForeSite SS
The ForeSite SS is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the ForeSite SS
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
ForeSite SS as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
M
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.24
0.1
0.37
0.37
0.74
1
1.17
1.17
2.34
10
3.69
3.69
7.38