8.1 FAULT JUDGMENT
If you come across any problem, you can check the
reasons.
FAULT
RESON
EXCLUDE
No display
Not connected to the power
supply;
Fuse is broken;
Power supply damaged;
Plug in the power line;
Replace the fuse;
Replacement of power transformer;
Contact factory
Weighing
unstable
Bad working conditions;
The wind screen is open
;
Something
between
the
table and balance
;
The power unstable
;
Weighing unstable
;
Keep the environment stable, close
the windows and doors
;
Close the glass door
;
The weighing
is wrong
The balance not calibrated.
No tare before weighing.
No well leveled.
Calibration.
Tare before weighing.
Adjust the level feet.
07
Ⅱ
SETTINGS
TABLE OF CONTENTS
1.0 KNOW YOUR BALANCE
2.0 SPECIFICATIONS
3.0 UNPACKING THE BALANCE
4.0 CALIBRATION
5.0 LOCATING THE BALANCE
6.0 SETTING UP THE BALANCE
6.1 ASSEMBLING THE BALANCE
6.2 LEVELLING THE BALANCE
7.0 OPERATION SETTINGS
7.1 SENSITIVITY SETTING
7.2 SPEED SETTING
7.3 WEIGHING UNIT SETTING
7.4 COUNTING FUNCTION
7.5 PERCENTAGE FUNCTION
7.6 DENSITY FUNCTION
8.1 FAULT JUDGMENT
9.1 DATA OUTPUT