background image

02

 

5.0 LOCATING THE BALANCE 

• The balance should not be placed in a location that will reduce the accuracy. 
• Avoid extremes of temperature. Do not place in direct sunlight or near air 
conditioning vents. 
• Avoid unsuitable tables. The table or floor must be rigid ,no vibration. 
•  Avoid  unstable  power  sources.  Do  not  use  near  large  users  of  electricity 
such as welding equipment or large motors. 
• Do not place near vibrating machinery. 
•  Avoid  high  humidity  that  might  cause  condensation.  Avoid  direct  contact 

with water. Do not spray or immerse the balances in water. 

• Avoid air movement such as from fans or opening doors. Do not place near 
open windows or air-conditioning vents. 
• Keep the balance clean. Do not stack material on the balances when they 
are not in use. 

 

6.0 SETTING UP THE BALANCE 

 

6.1 ASSEMBLING THE BALANCE 

• Locate balance on solid surface, free from vibration 
• Open the sliding door and gently place the stainless steel top. 
• Level balance using the adjustable feet and the bubble level 
• Connect the power to the balance 
• For best performance, let the balance warm up for 30-60 min. and calibrate 
before using 

6.2 LEVELLING THE BALANCE 

After placing the balance in a suitable place, level it by using the bubble level. 

To level the balance turn the two adjustable feet at the rear of the balance until 

the bubble in the bubble level is center. 

 

 

 

 

 

 

 

03

 

7.0 OPERATION SETTINGS 

 

 

 

 

 

 

 

 

 

 

 

 

      1.State Icon        2. Function State    3.Weighing Digits     

      4.Notes              5.Operation Keys 

 

 

7.1 SENSITIVITY SETTING   

--Highest Sensitivity   

-Normal Sensitivity   

-Sensitivity

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

04 

Reviews: