background image

15  

 

UM Termometri serie 7Vio ENG rev.0 14.04.2021

 

 

Other Features

 

 

 

Automatic switch-off

 

The instrument shuts down automatically after 20 minutes of inactivity. To disable automatic shutdown

 

To disable automatic shutdown, go to parameter P9.9 in the Setup menu and select Off.

 

 

 

Offset adjustment

 

All instruments are factory pre-calibrated and remain stable over time, but in the event of a reading error, 
the instrument can be recalibrated to eliminate this error.

 

To recalibrate the instrument at 0.0°C proceed as follows:

 

 

prepare ice with double-distilled water;

 

 

place the ice cubes in a 500ml beaker up to half its volume;

 

 

add double-distilled water until it reaches ¾ of its volume;

 

 

place the beaker on a magnetic stirrer under gentle agitation;

 

 

immerse the probe in the beaker, ensuring that it is immersed in water and does not touch the wall of 
the beaker;

 

 

cover the beaker with polystyrene to isolate the water/ice mixture from the air as much as possible;

 

 

stir for about 10 minutes;

 

 

in this way a temperature of 0°C is reached.

 

If the instrument reads 0.0°C (± measurement chain accuracy) then the instrument is operating correctly 
and no adjustment is required, if not, go to the Setup menu P7.2, adjust the value using the arrows and 

confirm with the key. 

 or 

 arrows and confirm with the key 

 .

 

In the centre of the display, the corrected temperature is displayed in large letters, while the partial value 
of the adjustment is displayed below it.

 

The adjustment affects the entire measurement scale, e.g. if you correct by -1.0°C, then the measurement 
shifts by -1.0°C over the whole range.

 

The  setting  made  in  this way remains  in  the  memory  both after switching  off  the  instrument  and when 
replacing the batteries.

 

If the probe is replaced, reset the factory calibration from the Setup menu P7.6, and proceed with the new 
calibration if necessary.

 

 

Note

: if there is no probe connected, it is not possible to change the offset.

 

A "CONNECT THE PROBE" error will be displayed, accompanied by a flashing icon  

.

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for 7 Vio

Page 1: ...USER MANUAL Temp7 PT100 Temp7 K T Temp7 NTC...

Page 2: ...1 UM Termometri serie 7Vio ENG rev 0 14 04 2021...

Page 3: ...Features 6 Datasheet 6 Instrument Description 7 Display 7 Keypad 7 LED 8 Setup 8 Supplied component 8 Start up 8 Switching On Off 8 Batteries replacement 8 Instrument transportation 9 Keys Function 9...

Page 4: ...serie 7Vio ENG rev 0 14 04 2021 Warranty period and limitations 16 Disposal of electrical devices 16 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax 39059653282 www x...

Page 5: ...ON for a medium risk hazardous situation which could lead to serious injury or death if not avoided ATTENTION for a dangerous situation with reduced risk which can cause material damage data loss or m...

Page 6: ...ime in adverse conditions exposure to direct light heat sources or places saturated by gas or vapours or in environments with conditions different from those mentioned in this manual Device maintenanc...

Page 7: ...curacy 2 C 200 100 C 0 2 C 99 9 199 9 C 2 C 200 999 C 0 1 of the read 0 4 C over 150 C 0 25 of the read 1 C below 150 C 0 2 C System Display High definition colours LCD Backlit Brightness setup Manual...

Page 8: ...uipment ESC key to go back at the measurement mode OK Setup key to confirm the chosen values and enter the Setup from the measurement mode DOWN HOLD key to scroll menus and values in the Setup Short p...

Page 9: ...witching On Off Turn on the system by pressing the button The display initially activates all segments and then appears model and firmware of the instrument settings related to the most important para...

Page 10: ...go back to measurement mode Short MAX MIN memory activation MAX value display and MIN value display It is used to scroll and change the parameters of the Setup menu Long Press 3 sec 3 seconds pressure...

Page 11: ...s and icons Symbols Description Type of probe connected Error in measurement Battery charge indication Measurement stability indicator The bars scroll if the measurement is not stable Press the direct...

Page 12: ...SETUP mode select the parameter you want to edit by using the directional keys and confirming with Within the selected menu move between the different programs using the directional buttons and press...

Page 13: ...e temperature probe connected If menu P 7 2 is accessed without a temperature sensor connected the display will show the error message CONNECT THE PROBE accompanied by the flashing icon Note For the T...

Page 14: ...n three different levels of display brightness LOW low NORMAL medium HIGH high Note Keeping the display bright always adversely affects battery life P9 5 Sleep mode Access this setup menu to select wh...

Page 15: ...ill flash MIN MAX REC Press the key once to see the maximum value reached up to that moment press it another time to see the minimum value reached up to that moment Pressing again will return the inst...

Page 16: ...isolate the water ice mixture from the air as much as possible stir for about 10 minutes in this way a temperature of 0 C is reached If the instrument reads 0 0 C measurement chain accuracy then the i...

Page 17: ...urer will repair or replace defective components This warranty is valid only and exclusively on the electronic parts of the device and does not apply if the product has been damaged used incorrectly e...

Reviews: