The UNESCO 1966b definition can also be used, which provides that the salinity in ppt is expressed with the
following formula:
S
ppt
=-0.08996+28.212.80832R
2
-10.67869R
3
+5.98624R
4
-1.32311R
5
Where R= Cond sample (at 15°) / 42.914 mS/cm (Conductivity of Copenhagen Seawater Standard)
Data Logger Function
This series of devices has the possibility of recording values in GLP format on the instrument's internal
memory.
•
The instrument can save up to 1000 data in total. Once the memory is finished, the values are NOT
overwritten. In measure mode, the number of data stored for that parameter appears next to the
M+
icon.
•
It is possible to recall and consult the values on the display or download them to a PC using the
appropriate software.
•
If you have the possibility to work directly connected to the PC, the data are automatically saved on
the software without having memory limitations.
•
Recordings can be acquired
manually
(MANUAL) or
automatically at preset frequencies
(HOURS -
MINUTES).
PC Connection
:
connect the USB cable inside each package to the USB port on the top panel of the
instrument and the other end to a COM port on the computer.
Use the USB cable supplied with the instrument only.
Data Logger on external printer:
Purchase the printer separately, connect it with RS232 cable to the connector identified with "Printer" on
the rear panel of the instrument, connect the power supply to the grid power and switch it on by pressing
the I / O
button. For further information, refer to the technical manual of the printer.
Use the printer specified by the manufacturer only. Contact your local distributor for purchasing or further
information.
The original printer is already equipped with a roll of plain paper, power supply and RS232 cable for
instrument connection.
•
Setup for Data Logger parameter
•
In measure mode press button
to access the SETUP menu.
•
Use the directional keys to move to
LOG SETTINGS P8.0
and access the menu by pressing the key
.
•
Move with the keys
and
to select the program to access.
The table below shows the setup menu structure for the Data Logger mode; for each program, there are
the options that the user can choose and the default value: