better the temperature correction, the lower the error. The only way to eliminate this error is not to use the
correction factor, acting directly on the sample temperature.
Select 0.00% / °C as the temperature coefficient to deactivate the compensation. The displayed Conductivity
value refers to the temperature value measured by the probe and not related to a reference temperature.
P3.6 COND calibration data
Access this menu to view (select “
VIEW
”) or print (select “
”) information on the last calibration
performed:
•
VIEW:
The following screens will scroll automatically on the display:
•
First screen: CALIBRATION DATE and TIME and beakers indicating the buffers used.
•
Second and possibly third, fourth and fifth screens: value of the actual cell constant in the
measurement range indicated by the beaker.
Note
: The instrument accepts calibrations with a maximum tolerance of 40% on the nominal
value of the cell constant only.
•
PRINT:
The following information are printed
•
Model and Serial Number of the instrument
•
Calibration Date and time
•
Measuring Range with relative cell constant applied
IMPORTANT:
Make sure that the printer is connected and turned on and that the paper roll and cartridge are
correctly positioned. For more information, refer to the printer user manual.
NOTE:
Use the original printer specified by the manufacturer only. For information, contact your local distributor.
P3.7 COND Calibration expiration
Access this menu to set a calibration deadline; this option is fundamental in GLP protocols.
•
By default, no calibration deadline is set. Use the directional keys to select HOURS or DAYS and access
with
. Use the directional keys to change the number that appears in the centre of the display,
entering the hours or days that must elapse between two settings, and confirm with key
.
•
If a calibration deadline is set, the icon is present on the display in measure mode
.
When the calibration deadline is activated, the instrument prevents further measurements.
The error icon
and the icon representing the calibration deadline
flash on the display.
The string “
MAKE A NEW CAL
” invites the user to perform a new calibration of the pH sensor to be
able to work again.
Press the key
to start the calibration.
P3.8 COND Reset parameter