25
Rev. 1.0 31/01/2020
EN
•
ORP Parameter Setup
•
In measure mode press the key
to access the SETUP menu.
•
Use the directional keys to move to
ORP SETTINGS P2.0
and access the menu by pressing the key
.
•
Move with the keys
and
to select the program to access.
The table below shows the Setup menu structure for the ORP parameter; for each program there are the
options that the user can choose and the default value:
•
Composition of the setup menu for the ORP parameter
Program
Description
Options
Factory Default Settings
P2.6
CALIBRATION DATA
-
-
P2.7
SET DUE CAL
NO – HOURS - DAYS
NO
P2.8
RESET SETTINGS
YES – NO
NO
P2.9
TEMPERATURE CAL
YES – NO
-
P2.6 Calibration Data
Access this menu to view (select “
VIEW
”) or print (select “
”) information on the last calibration
performed:
•
VIEW:
The following screens will scroll automatically on the display:
•
First screen: CALIBRATION DATE and TIME.
•
Second screen: OFFSET value of the electrode expressed in mV.
•
Third screen: TEMPERATURE at which the calibration was performed.
•
PRINT:
The following information are printed:
•
Model and Serial Number of the instrument.
•
Calibration Date and Time
•
OFFSET value expressed in mV
•
Calibration TEMPERATURE
IMPORTANT:
Make sure that the printer is connected, turned on and that the paper roll and cartridge are correctly
positioned. For more information, refer to the printer user manual.
NOTE:
Use the original printer indicated by the manufacturer only. For information, contact your local distributor.
P2.7 ORP calibration deadline
Access this menu to set a calibration deadline; this option is fundamental in GLP protocols.