Triggered mode - Burst of frames
For more information please see:
Frame Burst Start
In this mode each trigger pulse triggers defined number of exposed frames.
Figure 68: triggered burst of frames – frame burst start
Frame Burst Active
If trigger is level-sensitive, it can be used to control image acquisition.
Figure 69: triggered burst of frames – frame burst active
Triggered mode - Exposure defined by trigger pulse length
In this mode the exposure is defined by trigger pulse length. This can be used to achieve longer exposure than allowed by
. Also, it can
be used to trigger several images in sequence with different exposure time. Exposure time is measured and reported in image metadata.
Please see:
Exposure Defined by Trigger Pulse Length
Figure 70: exposure defined by trigger pulse length
XIMEA Technical Manual, Version: v240108