14
SCATIMEON
Assertion time (in clock periods) of the SCA real-
time digital outputs
15
SCATIMEOFF
Minimum off time (in clock periods) of the SCA
real-time digital outputs
16
SCA0LIMLO
Lower limit for SCA0
17
SCA0LIMHI
Upper limit for SCA0
18
SCA1LIMLO
Lower limit for SCA1
19
SCA1LIMHI
Upper limit for SCA1
20
SCA2LIMLO
Lower limit for SCA2
21
SCA2LIMHI
Upper limit for SCA2
22
SCA3LIMLO
Lower limit for SCA3
23
SCA3LIMHI
Upper limit for SCA3
Appendix D. PARSET Specification
Twenty-four PARSET tables each correspond to a
Peaking Time
, i.e. these are all the
parameters that are stored and retrieved for each
Peaking Time
. See §2.5 for details.
ID
Parameter
Description
0
NUMPARSET
Number of PARSET parameters NOT including
NUMPARSET and PARVERSION (used
internally)
1
PARVERSION
PARSET version number (used internally)
2
FASTLEN
Fast filter ramp length
3
FASTGAP
Fast filter gap
4
FSCALE
Controls FiPPI fast filter scaling
5
MINWIDTH
Minimum time above threshold for fast filter
6
MAXWIDTH
Maximum time above threshold for fast filter
(detects fast filter pileup)
7
SLOWLEN
Slow Filter Ramp length
8
SLOWGAP
Slow Filter gap
9
PEAKMODE
Peak Sensing (0) or Peak Sampling (1)
10
PEAKINT
Minimum spacing between pulses
11
PEAKSAM
Energy Sampling point
12
BFACTOR
Ratio between slow and intermediate filters
13
BLFILTER
Baseline filter average length = 32768/BLFILTER
14
THRESHOLD
Fast filter threshold corresponding to the current
GENSET.
15
BASETHRESH
Intermediate filter threshold corresponding to the
current GENSET. Needs to be set properly to get
good baselines.
16
SLOWTHRESH
Energy filter threshold corresponding to the
current GENSET (only needed for light elements
– normally set to 0 to disable)
17-21
GAINTWEAK0-4
If non-zero, this per-GENSET parameter modifies
the GAINBASE setting in the GENSET to arrive
at the GAINDAC setting. E.g. If GENSET 3 is in
use, GAINDAC = GA GAINTWEAK
3.