Table 4 

Recommended Separation Distance 

This device can be used under the environment that radiated RF disturbances are controlled. User should maintain a minimum 

distance between portable and mobile RF communications equipment to prevent electromagnetic interference. Following 

recommended distance is calculated according to the maximum output power of the communication equipment. 

Rated maximum output 

power of transmitter (W) 

Separation distance according to frequency of transmitter (m) 

150kHz -2MHz 

P

d

2

.

1

=

 

80MHz -800MHz 

P

d

2

.

1

=

 

800MHz -2.5GHz 

P

d

3

.

2

=

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance “d” in meters can 

be estimated using the equation applicable to the frequency of transmitter, where “p” is the maximum output power rating of 

the transmitter in watts according to the transmitter manufacturer. 

Note1: at 80M~800MHz, the separation distance for the higher frequency range applies. 

Note2: these guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and refection 

from structures, objects and people. 

 

 

2. Overviews 

2.1

 

Indication for Use 

XFT-2001EA Foot Drop System is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to 

upper motor neurons or pathways to the spinal cord. During the swing phase of walking, the XFT-2001EA electrically stimulates 

the appropriate muscles that cause ankle dorsiflexion and may thus improve the individual’s gait. Medical benefits of Functional 

Electrical Stimulation (FES) may include prevention/retardation of disuse atrophy, increased local blood flow, muscle 

re-education, and maintained or increased joint range of motion. 

 

How does XFT-2001EA Foot Drop System work? 

When the leg swings to the angle of threshold, the electrical stimulation will be triggered. 

 

2.3 Features and Innovation 

 

Bluetooth 4.0, more stable and reliable data transmission, easy operation and setting. 

 

Thin and light Stim Unit & Cuff with patented design, easy and simple to wear. 

 

Integrated stainless steel electrode, fully enclosed waterproof design, no need for additional consumables, easy to clean and 

store. 

 

Automatic shutdown function: low voltage or automatic shutdown after 30 minutes of standby. 

 

Built-in rechargeable battery; one single charge for long-term use. 

 

Electrical stimulation constant current output & low voltage alarm function. 

 

It has two modes of walking and training, which can be selected according to the patient's state; 

 

Equipped with APP management software; able to manage user information and store gait data. It is simple and convenient to 

set the parameters of the Stim Unit through the mobile phone. 

Summary of Contents for G4

Page 1: ...Foot Drop System User Manual Caution Federal law restricts this device to sale by or on the order of a practitioner licensed by the law of the State in which he she practices to use or order the use...

Page 2: ...and Safety 3 2 Overviews 8 3 Product Illustration 9 4 Operation Instruction 12 5 Attentions 17 6 Care and Maintenance 17 7 FAQ Troubleshooting 18 8 Product Configuration 18 9 Product classification 19...

Page 3: ...the device battery accessory packing in accordance with the legal obligation in your area Contraindications Powered muscle stimulators should not be used on patients with cardiac demand pacemakers Do...

Page 4: ...ds 10mA cm2 r m s Please use with caution if the used areas have structural deformity This product should be conducted by doctors Patients should keep stable and not move while using this unit Patient...

Page 5: ...ctions may put the user at undue risk of injury Adverse Reactions Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators Electromagnetic Compati...

Page 6: ...is used in such an environment Immunity test IEC60601 test level Compliance Guidance Electrostatic discharge ESD IEC 61000 4 2 6kV contact 8kV air 6kV contact 8kV air Floors should be wood concrete or...

Page 7: ...ermined by an electromagnetic site survey a should be less than the compliance level in each frequency range b interference may occur in the vicinity of equipment marked with the following symbol Radi...

Page 8: ...s the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord During the swing phase of walking the XFT 2001EA electrically stimulates th...

Page 9: ...for 1 4 hours a day every morning and evening 20 minutes each time 3rd week later Walk for 4 8 hours a day every morning and evening 20 minutes each time Note take off the cuff for 15 minutes after e...

Page 10: ...working states of the unit such as gait mode training mode electrode loose low battery icon electrical stimulation output icon electrical stimulation intensity etc 3 2 2 Indicators Power on indicatio...

Page 11: ...is low there will be a battery icon display battery display zero capacity at low voltage and flash at 1 second interval When the battery voltage is as low as the shutdown voltage the unit automaticall...

Page 12: ...When the Stim Unit is in the pause state press the Electrical Stimulation Intensity Adjustment Button to activate the electrical stimulation intensity press the up button to increase the electrical st...

Page 13: ...stool relax the affected leg and bend naturally Place the Stim Unit to correct position under the knee 4 1 3 3 Power on and operate Turn on Press and hold the switch mode composite button for 2 secon...

Page 14: ...it receives a reply from the Stim Unit and pauses the walk Continue The App sends the gait mode start at which time the current intensity given in the parameter setting is the current intensity before...

Page 15: ...lation stop the electrical stimulation by the release and continue to hold the electrical stimulation for continuous output The parameters that can be adjusted are the electrical stimulation intensity...

Page 16: ...pp sends an evaluation mode to start the command to the Stim Unit receives the Stim Unit reply and enters the evaluation work interface Stop After long press and stop for 2s send the evaluation mode s...

Page 17: ...in Care Before and after using the device please examine the skin it is normal that skin slightly turns red which shows faster blood flow Allergy Prevention Advice Do not position on the skin with mak...

Page 18: ...ng to use the device Remember to regularly ventilate the skin covered by the Stim Unit Q4 The unit display battery icon is turned off after 1 second interval flashing A This indicates that the Stim Un...

Page 19: ...ed 8 5 Product configuration Stim Unit 1pc Power Adapter 1set User manual 1pc 9 Product classification a Classified by type of electric shock internal power supply b The application part is classified...

Page 20: ...epaired by unauthorized repair store 3 When asking for warranty service please take with the warranty card It is charged according to the stipulation of the repair service of the warranty Please conta...

Page 21: ...trasse 80 D 20537 Hamburg Germany US Agent Yang LingYun 605 Sherwood Forest Place Cary North Carolina Cary NC 27519 Phone 919 4751044 Ext Email Usagency 126 Com Date 2018 12 17 Doc No XFT 2001EA E Ver...

Reviews: