XFT Medical G4 User Manual Download Page 17

5. Attentions 

5.1 Device Power Off 

 

Force shut down: Press the shutdown button. 
System shut down:   
1) When the treatment duration set by the channel has ended, the functions stop. 
2) When the battery voltage is low, the system will automatically shut down. 
3) The Stim Unit will automatically shut down when it is in standby (no electrical stimulation output) for 2 hours. 
 

5.2 Troubleshooting

 

Malfunction indicator will show the following troubles: 

 

5.2.1

 

Electrodes loose

 

The malfunction indicator flashes slowly (the frequency is 1 Hz). When the system detects that the electrode is off, it indicator 
will display and the device will stop running. This trouble can be resolved by simply connecting the electrode pads. 
 
5.2.2 Low battery voltage 
When the battery voltage is low, there will be a battery icon display (the battery display capacity is empty) at low voltage and 
flash at 1 second intervals. When the battery voltage is as low as the shutdown voltage, the unit automatically shuts down. 
 

5.3 Integrated electrode use precautions 

The integrated electrode piece can be replaced and be used for long-term. Pay attention to clean the contact surface or pay 
attention to clean the skin surface during use (users with more hair should pay attention to trimming, and conductive adhesive can 
also be used to increase conductivity). 
 

5.4

 

Skin Care 

Before and after using the device, please examine the skin, it is normal that skin slightly turns red, which shows faster blood flow. 

 

Allergy Prevention Advice: 

 

Do not position on the skin with makeup or oil. 

 

Remove arm’s hair for better electrical conductivity. Electric razor or a pair of scissors is recommended. 

 

If any skin irritation or allergy occurs, please stop using the device immediately and follow doctor’s instructions. 

 

Do not position on allergic area. 

 

6. Care and Maintenance 

6.1 Device Storage 

 

Please do not store it in the place of direct sunlight, high temperature, moist, dusty, or corrosive gas. 

 

Please store it in the place where children cannot reach. 

 

The user does not need to maintain the hose device, please ask the seller or manufacturer. 

 

Please use wet cloth with neutral detergent or alcohol to clean the surface of the Stim Unit. 

 

Please do not immerse the electronic components into to water. 

 

Please do not throw, tread on, or heavy press the device. 

 

6.2 Product Life 

The product is designed to be used for 5 year, after which, please deal with product and battery according to the provisions of the 

electronic products processing. 

 

6.3

 

Use of Rechargeable Battery 

 

Do not charge/discharge exceed the specified current.   

 

Do not short the battery/battery pack. May cause permanent damage to the battery/battery. 

 

Do not burn or destroy the battery/battery pack. 

 

Do not expose the device to adverse conditions such as extreme temperatures, deep cycling and overcharging, otherwise 

battery life may be reduced. 

Summary of Contents for G4

Page 1: ...Foot Drop System User Manual Caution Federal law restricts this device to sale by or on the order of a practitioner licensed by the law of the State in which he she practices to use or order the use...

Page 2: ...and Safety 3 2 Overviews 8 3 Product Illustration 9 4 Operation Instruction 12 5 Attentions 17 6 Care and Maintenance 17 7 FAQ Troubleshooting 18 8 Product Configuration 18 9 Product classification 19...

Page 3: ...the device battery accessory packing in accordance with the legal obligation in your area Contraindications Powered muscle stimulators should not be used on patients with cardiac demand pacemakers Do...

Page 4: ...ds 10mA cm2 r m s Please use with caution if the used areas have structural deformity This product should be conducted by doctors Patients should keep stable and not move while using this unit Patient...

Page 5: ...ctions may put the user at undue risk of injury Adverse Reactions Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators Electromagnetic Compati...

Page 6: ...is used in such an environment Immunity test IEC60601 test level Compliance Guidance Electrostatic discharge ESD IEC 61000 4 2 6kV contact 8kV air 6kV contact 8kV air Floors should be wood concrete or...

Page 7: ...ermined by an electromagnetic site survey a should be less than the compliance level in each frequency range b interference may occur in the vicinity of equipment marked with the following symbol Radi...

Page 8: ...s the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord During the swing phase of walking the XFT 2001EA electrically stimulates th...

Page 9: ...for 1 4 hours a day every morning and evening 20 minutes each time 3rd week later Walk for 4 8 hours a day every morning and evening 20 minutes each time Note take off the cuff for 15 minutes after e...

Page 10: ...working states of the unit such as gait mode training mode electrode loose low battery icon electrical stimulation output icon electrical stimulation intensity etc 3 2 2 Indicators Power on indicatio...

Page 11: ...is low there will be a battery icon display battery display zero capacity at low voltage and flash at 1 second interval When the battery voltage is as low as the shutdown voltage the unit automaticall...

Page 12: ...When the Stim Unit is in the pause state press the Electrical Stimulation Intensity Adjustment Button to activate the electrical stimulation intensity press the up button to increase the electrical st...

Page 13: ...stool relax the affected leg and bend naturally Place the Stim Unit to correct position under the knee 4 1 3 3 Power on and operate Turn on Press and hold the switch mode composite button for 2 secon...

Page 14: ...it receives a reply from the Stim Unit and pauses the walk Continue The App sends the gait mode start at which time the current intensity given in the parameter setting is the current intensity before...

Page 15: ...lation stop the electrical stimulation by the release and continue to hold the electrical stimulation for continuous output The parameters that can be adjusted are the electrical stimulation intensity...

Page 16: ...pp sends an evaluation mode to start the command to the Stim Unit receives the Stim Unit reply and enters the evaluation work interface Stop After long press and stop for 2s send the evaluation mode s...

Page 17: ...in Care Before and after using the device please examine the skin it is normal that skin slightly turns red which shows faster blood flow Allergy Prevention Advice Do not position on the skin with mak...

Page 18: ...ng to use the device Remember to regularly ventilate the skin covered by the Stim Unit Q4 The unit display battery icon is turned off after 1 second interval flashing A This indicates that the Stim Un...

Page 19: ...ed 8 5 Product configuration Stim Unit 1pc Power Adapter 1set User manual 1pc 9 Product classification a Classified by type of electric shock internal power supply b The application part is classified...

Page 20: ...epaired by unauthorized repair store 3 When asking for warranty service please take with the warranty card It is charged according to the stipulation of the repair service of the warranty Please conta...

Page 21: ...trasse 80 D 20537 Hamburg Germany US Agent Yang LingYun 605 Sherwood Forest Place Cary North Carolina Cary NC 27519 Phone 919 4751044 Ext Email Usagency 126 Com Date 2018 12 17 Doc No XFT 2001EA E Ver...

Reviews: