Status Registers
Overview
76
Operating Manual for Multichannel Functionality (GPIB-M)
Table 4.7
QUEStionable Status Register
VOLTage
Sub-Register
This shows whether the present voltage level is over or under the specified trip limit.
Table 4.8
VOLTage Sub-Register
Bit
Bit Weight
Bit Name
Description
0
1
VOLTage Summary
Reflects a summary of the
VOLTage Sub-Register.
1
2
CURRent Summary
Reflects a summary of the
CURRent Sub-Register.
2
4
TIME Not
implemented
3
8
Reserved
4
16
Over Temperature
Indicates the OTP circuit has
tripped.
5
32
FREQuency Summary
Not implemented
6
64
PHASe Summary
Not implemented
7
128
MODulation Summary
Not implemented
8
256
CALibration
Indicates an error in the unit
calibration.
9
512
Not Used
Not implemented
10
1024
Not Used
Not implemented
11
2048
AC Off
Indicates an AC Supply failure.
12
4096
UNRegulated
Indicates that the output is not
regulated in either Constant
Voltage mode, Constant Current
mode or Constant Power mode.
Reflects the inverse of the
Operation Regulation Summary bit.
13
8192
INSTrument Summary
Not implemented
14
16384
Command Warning
Not implemented
15
32768
Not Used
Always zero
Bit
Bit Weight
Bit Name
Description
0
1
Over VOLtage
Set if the supply’s output voltage
exceeds the over-voltage trip level,
either user-specified variable trip limit,
or the fixed trip limit.
1
2
Under VOLtage
Set if the supply’s output voltage is less
than the user-specified under-voltage
trip level (variable trip limit) and the
supply is in Operation state.
Summary of Contents for GPIB-M-HPD
Page 2: ......
Page 18: ...List of Figures xvi Operating Manual for Multichannel Functionality GPIB M...
Page 94: ...Current Sharing Operation 92 Operating Manual for Multichannel Functionality GPIB M...
Page 98: ...GPIB Performance Specifications 96 Operating Manual for Multichannel Functionality GPIB M...
Page 118: ...SCPI Command Reference Expressions 116 Operating Manual for Multichannel Functionality GPIB M...
Page 133: ......