Installation and Configuration
Configure for GPIB Operation
34
Operating Manual for Multichannel Functionality (GPIB-M)
Configure for GPIB Operation
The power supply is ready for GPIB communication when it leaves the factory, but
you may change any of the control settings. The default remote control setting is
GPIB, address 2.
Change
Remote
Control
Source
Select an interface for remote control.
SYST:REM:SOUR [GPIB|MCHannel]
where
•
GPIB:
control via GPIB interface
•
MCHannel:
multichannel operation using the CANbus interface
Set GPIB
Address
Change the GPIB address.
SYST:COMM:GPIB:ADDR <GPIB-address>
where
<GPIB-address> may be a value from 1 to 30.
Be sure to make a note of the GPIB address. If the address is unknown, use a feature
such as “Scan for Instruments” on National Instruments’ Measurement and
Automation Explorer.