U
SER
M
ANUAL
C
OMMISSIONING
30
Continuous current of the DC power supply
A 12 V DC power supply must deliver a continuous
current of 1 A with 3 A peaks. Use a regulated power
supply or a car battery and note the required voltage and
current values.
Please note the following sequence:
Connect the power cable to port
(L)
on the X
ORAYA
Z7.
(
Connections and controls)
Tighten the connector screws.
Connect the black plug of the cable to 0 V or ground.
Connect the red plug to the power supply.
The X
ORAYA
Z7 is securely connected to the measuring environment.
Once the supply voltage is established, the X
ORAYA
Z7 turns on and displays
its operational status via the power button LED
(K)
.
CAUTION
Damage due to improper device shutdown
Disconnecting the power supply during operation may
cause data loss and the destruction of the X
ORAYA
Z7.
Never disconnect the X
ORAYA
Z7 from the power
supply while in operation.
Ensure proper connector seating and tighten the
screws if possible.
Only shut down the X
ORAYA
Z7 through the
X
ORAYA
Suite or the power button on the front panel.
Pull the black plug connected to ground last when
disconnecting the X
ORAYA
Z7 from the power supply.