I N S T R U M E N T C A L I B R A T I O N
3-3
3.3
Calibration Procedure
A calibration procedure consists of a white measurement
followed by a black measurement. The instrument notifies you
when a calibration is required.
NOTE:
The instrument must be calibrated with the target window
removed when using the instrument with the shoe extended
(unlatched).
To perform a calibration:
1.
At the Calibration menu screen, position the target window
over the white ceramic disk as previously explained.
C A L I B R A T I O N
<
M e a s u r e W h i t e R e f
>
S t a t u s : C a l t i m e u p
S / N : * * * * * *
A p S i z e : 8 . 0 m m
2.
Press the instrument firmly to the shoe. Hold steady until
the screen indicates the white calibration is completed.
Release instrument when <
Success!
> is displayed.
3.
Position the target window over the black port opening as
previously explained.
4.
Press the instrument firmly to the shoe. Hold steady until
the screen indicates the black calibration is completed.
5.
Store the calibration reference in a dry, dust free area,
away from direct exposure to light.