D T P 3 2 H S D E N S I T O M E T E R
6
Strip Positioning and Patch Requirements
The strip width must extend a minimum of 0.75” (19mm) to the left of the alignment mark
(optical path) when measuring. This ensures the read switch is activated when the strip is
inserted into the instrument.
TOP VIEW
0.75"
(19mm)
min.
4" (101.6mm) max. to paper stop
Paper Stop
Alignment Mark
0.50" (12.7mm) min. patch width
1.2" (30.5mm)
min. leader
length
0.40" (10.2mm)
min. patch
height