D T P 3 2 H S D E N S I T O M E T E R
5
Instrument Location Requirements
To ensure reliable read switch activation, the instrument must sit on a reasonably flat, level,
and stable surface.
Strip Feeding Requirements
To ensure consistent strip readings, you must feed the strip into the instrument at a
continual pace (without pausing) and speed before it is taken in by the drive motor. More
specifically, a strip cannot be fed into the instrument too quickly (more than 7.2
inches/second or 183 mm/second), too slowly (less than 2.6 inches/second or 66
mm/second), or intermittently (pausing one or more times). If you are experiencing any
strip-feeding problems, observe how fast the motor is moving the strip (about 5.4
inches/second or 137 mm/second) and try to match the speed as closely as possible.
Before feeding a second strip or a second pass of a multi-pass strip into the instrument, you
must momentarily pause (approx. 3 seconds) to allow the instrument to process data from
the first measurement. The LED is not lit when the instrument is processing data.
Instrument LED Indicator
The LED indicates a variety of instrument operation conditions, such as calibration mode
and operation. Below is a complete list of conditions reported by the LED.
Operation Mode
•
Solid Green
—self-test passed and instrument is ready for use with defined strip.
•
Slow Flashing Green—
strip reading in progress (strip being read) or ready for next pass
(multi-pass strip definition)
.
•
Fast Flashing Green
—strip reading was attempted but the wrong strip was read, or strip
was skewed during reading.
•
Slow Flashing Amber/Green
—a configuration barcode was scanned and the instrument is
waiting for the 2
nd
or 3
rd
barcode pass.
•
Fast Flashing Amber/Green
—there was an error in recognizing a barcode or the
configuration failed. Reread the entire barcode, starting with the first pass.
Calibration Mode
•
Slow Flashing Green
—instrument calibration in progress (strip being read).
•
Solid Green after Flashing Green
—instrument calibration was successful and instrument
is ready for use.
•
Fast Flashing Green
—instrument calibration failed.
Error/Reset Mode
NOTE:
When the instrument is first turned on, the motor runs, and the LED illuminates red while
performing self-tests. Self-tests are completed in approximately 2 to 3 seconds, at which time the LED
turns green and the motor stops under normal conditions.
•
Fast Flashing Red
—incorrect power supply attached. Verify correct 12-volt power supply
is used.
•
Slow Flashing Red
—a factory reset is in progress.
•
Flashing Red/Green
— the host firmware loader program is reprogramming the
instrument.
Summary of Contents for DC240 - DocuColor 240 Color Laser
Page 1: ...DTP32HS Auto Scanning Densitometer Operator s Manual ...
Page 2: ......
Page 20: ...Baud Rate 1200 ...
Page 21: ...Baud Rate 2400 ...
Page 22: ...Baud Rate 4800 ...
Page 23: ...Baud Rate 9600 ...
Page 24: ...Baud Rate 19200 ...
Page 26: ...Auto Transmit On ...
Page 27: ...Auto Transmit Off ...
Page 29: ...Decimal Point On ...
Page 30: ...Decimal Point Off ...
Page 32: ...Times Ten On ...
Page 33: ...Times Ten Off ...
Page 35: ...Data After Pass On ...
Page 36: ...Data After Pass Off ...
Page 38: ...Delimiter CR ...
Page 39: ...Delimiter CR LF ...
Page 41: ...Separator Space ...
Page 42: ...Separator Comma ...
Page 43: ...Separator Tab ...
Page 44: ...Separator CR ...
Page 45: ...Separator CR LF ...
Page 47: ...Min Max Off ...
Page 48: ...Min Max Min ...
Page 49: ...Min Max Max ...
Page 50: ...Min Max Min Max ...
Page 51: ......